DOI 10.1111/j.1747-0803.2006.00042.x
Медицинская библиотека · Congenital heart disease · 2006
Severe tricuspid stenosis during percutaneous occlusion of perimembranous ventricular septal defect with the new Amplatzer device.
СтатьяПеревод в очереди
Аннотация
BACKGROUND
The new Amplatzer device has been used with success for occlusion of the perimembranous ventricular septal defects and the index of complications is very low.
RESULTS
The reported case corresponds to a very rare severe acute tricuspid stenosis, caused by adherence of the prosthesis to the anterior tricuspid leaflet.
CONCLUSION
Device release was achieved by manipulation on the tricuspid apparatus.
Для цитирования: Christiani Luiz A, Bergman Fábio, Tress João C, Vanzillotta Pedro P, Pedra Carlos A C. Severe tricuspid stenosis during percutaneous occlusion of perimembranous ventricular septal defect with the new Amplatzer device.. Congenital heart disease. 2006. DOI: 10.1111/j.1747-0803.2006.00042.x.