Reproducibility of straylight measurement by C-Quant for assessment of retinal straylight using the compensation comparison method.

Guber Ivo, Bachmann Lucas M, Guber Josef, Bochmann Frank, Lange Alex P, Thiel Michael A

Клиническое исследованиеСравнительное исследованиеСтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Straylight gives the appearance of a veil of light thrown over a person's retinal image when there is a strong light source present. We examined the reproducibility of the measurements by C-Quant, and assessed its correlation to characteristics of the eye and subjects' age.

PARTICIPANTS AND

METHODS

Five repeated straylight measurements were taken using the dominant eye of 45 healthy subjects (age 21-59) with a BCVA of 20/20: 14 emmetropic, 16 myopic, eight hyperopic and seven with astigmatism. We assessed the extent of reproducibility of straylight measures using the intraclass correlation coefficient.

RESULTS

The mean straylight value of all measurements was 1.01 (SD 0.23, median 0.97, interquartile range 0.85-1.1). Per 10 years of age, straylight increased in average by 0.10 (95%CI 0.04 to 0.16, p < 0.01]. We found no independent association of refraction (range -5.25 dpt to +2 dpt) on straylight values (0.001; 95%CI -0.022 to 0.024, p = 0.92). Compared to emmetropic subjects, myopia reduced straylight (-.011; -0.024 to 0.02, p = 0.11), whereas higher straylight values (0.09; -0.01 to 0.20, p = 0.09) were observed in subjects with blue irises as compared to dark-colored irises when correcting for age. The intraclass correlation coefficient (ICC) of repeated measurements was 0.83 (95%CI 0.76 to 0.90).

CONCLUSIONS

Our study showed that straylight measurements with the C-Quant had a high reproducibility, i.e. a lack of large intra-observer variability, making it appropriate to be applied in long-term follow-up studies assessing the long-term effect of surgical procedures on the quality of vision.

Для цитирования: Guber Ivo, Bachmann Lucas M, Guber Josef, Bochmann Frank, Lange Alex P, Thiel Michael A. Reproducibility of straylight measurement by C-Quant for assessment of retinal straylight using the compensation comparison method.. Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie. 2011. DOI: 10.1007/s00417-011-1704-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.