DOI 10.1097/opx.0b013e31821bbd68

Медицинская библиотека · Optometry and vision science : official publication of the American Academy of Optometry · 2011

Clinical validation of point-source corneal topography in keratoplasty.

Vrijling Anne C L, Braaf Boy, Snellenburg Joris J, de Lange Fleur, Zaal Michel J W, van der Heijde Rob G L, Sicam Victor A D P

Сравнительное исследованиеСтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To validate the clinical performance of point-source corneal topography (PCT) in postpenetrating keratoplasty (PKP) eyes and to compare it with conventional Placido-based topography.

METHODS

Corneal elevation maps of the anterior corneal surface were obtained from 20 post-PKP corneas using PCT (VU topographer, prototype; VU University Medical Center, Amsterdam, The Netherlands) and Placido-based topography (Keratron, Optikon 2000, Rome, Italy). Corneal surface parameters are calculated in terms of radius and asphericity. Corneal aberrations were characterized using standard Zernike convention. An artificial surface with quadrafoil feature (SUMIPRO, Almelo, The Netherlands) was measured and used as a reference to assess instrument performance compared with the gold standard.

RESULTS

The differences (mean ± std of PCT - Placido) found between the two types of topographers in measurements of post-PKP eyes are 0.02 ± 0.21 mm (p=0.64) for radius of curvature, 0.14 ± 0.49 (p=0.23) for asphericity, -0.19 ± 1.67 μm (p=0.61) for corneal astigmatism, -0.25 ± 1.34 μm (p=0.41) for corneal coma, 0.23 ± 0.82 μm (p=0.23) for corneal trefoil, and 0.15 ± 0.28 μm (p=0.02) for corneal quadrafoil. The PCT measured the artificial surface more accurate (rms error 0.16 μm; 0.12 eq. Dpt.) than the Placido-based topographer (rms error 1.50 μm; 1.15 eq. Dpt.).

CONCLUSIONS

PCT is more accurate than Placido-based topography in measuring quadrafoil aberration.

Для цитирования: Vrijling Anne C L, Braaf Boy, Snellenburg Joris J, de Lange Fleur, Zaal Michel J W, van der Heijde Rob G L, Sicam Victor A D P. Clinical validation of point-source corneal topography in keratoplasty.. Optometry and vision science : official publication of the American Academy of Optometry. 2011. DOI: 10.1097/opx.0b013e31821bbd68.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.