de Vries Niels E, Webers Carroll A B, Touwslager Wouter R H, Bauer Noel J C, de Brabander John, Berendschot Tos T, Nuijts Rudy M M A
Аннотация
PURPOSE: To analyze the symptoms, etiology, and treatment of patient dissatisfaction after multifocal intraocular lens (IOL) implantation.
SETTING: Department of Ophthalmology, Maastricht University Medical Center, The Netherlands.
DESIGN: Case series.
METHODS: In this retrospective chart review, the main outcome measures were type of complaints, uncorrected and corrected distance visual acuities, uncorrected and distance-corrected near visual acuities, refractive state, pupil diameter and wavefront aberrometry measurements, and type of treatment.
RESULTS: Seventy-six eyes of 49 patients were included. Blurred vision (with or without photic phenomenon) was reported in 72 eyes (94.7%) and photic phenomena (with or without blurred vision) in 29 eyes (38.2%). Both symptoms were present in 25 eyes (32.9%). Residual ametropia and astigmatism, posterior capsule opacification, and a large pupil were the 3 most significant etiologies. Sixty-four eyes (84.2%) were amenable to therapy, with refractive surgery, spectacles, and laser capsulotomy the most frequent treatment modalities. Intraocular lens exchange was performed in 3 cases (4.0%).
CONCLUSION: The cause of dissatisfaction after implantation of a multifocal IOL can be identified and effective treatment measures taken in most cases.
SETTING: Department of Ophthalmology, Maastricht University Medical Center, The Netherlands.
DESIGN: Case series.
METHODS: In this retrospective chart review, the main outcome measures were type of complaints, uncorrected and corrected distance visual acuities, uncorrected and distance-corrected near visual acuities, refractive state, pupil diameter and wavefront aberrometry measurements, and type of treatment.
RESULTS: Seventy-six eyes of 49 patients were included. Blurred vision (with or without photic phenomenon) was reported in 72 eyes (94.7%) and photic phenomena (with or without blurred vision) in 29 eyes (38.2%). Both symptoms were present in 25 eyes (32.9%). Residual ametropia and astigmatism, posterior capsule opacification, and a large pupil were the 3 most significant etiologies. Sixty-four eyes (84.2%) were amenable to therapy, with refractive surgery, spectacles, and laser capsulotomy the most frequent treatment modalities. Intraocular lens exchange was performed in 3 cases (4.0%).
CONCLUSION: The cause of dissatisfaction after implantation of a multifocal IOL can be identified and effective treatment measures taken in most cases.