Iteratively re-weighted bi-cubic spline representation of corneal topography and its comparison to the standard methods.

Zhu Zhongxia, Janunts Edgar, Eppig Timo, Sauer Tomas, Langenbucher Achim

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: The aim of this study is to represent the corneal anterior surface by utilizing radius and height data extracted from a TMS-2N topographic system with three different mathematical approaches and to simulate the visual performance.

METHODS

An iteratively re-weighted bi-cubic spline method is introduced for the local representation of the corneal surface. For comparison, two standard mathematical global representation approaches are used: the general quadratic function and the higher order Taylor polynomial approach. First, these methods were applied in simulations using three corneal models. Then, two real eye examples were investigated: one eye with regular astigmatism, and one eye which had undergone refractive surgery. A ray-tracing program was developed to evaluate the imaging performance of these examples with each surface representation strategy at the best focus plane. A 6 mm pupil size was chosen for the simulation.

RESULTS

The fitting error (deviation) of the presented methods was compared. It was found that the accuracy of the topography representation was worst using the quadratic function and best with bicubic spline. The quadratic function cannot precisely describe the irregular corneal shape. In order to achieve a sub-micron fitting precision, the Taylor polynomial's order selection behaves adaptive to the corneal shape. The bi-cubic spline shows more stable performance. Considering the visual performance, the more precise the cornea representation is, the worse the visual performance is.

CONCLUSIONS

The re-weighted bi-cubic spline method is a reasonable and stable method for representing the anterior corneal surface in measurements using a Placido-ring-pattern-based corneal topographer.

Для цитирования: Zhu Zhongxia, Janunts Edgar, Eppig Timo, Sauer Tomas, Langenbucher Achim. Iteratively re-weighted bi-cubic spline representation of corneal topography and its comparison to the standard methods.. Zeitschrift fur medizinische Physik. 2010. DOI: 10.1016/j.zemedi.2010.07.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.