DOI 10.1016/j.jcrs.2010.05.018
Rotational stability of a single-piece toric acrylic intraocular lens.
Аннотация
PURPOSE: To assess the rotational stability of a single-piece toric hydrophobic acrylic intraocular lens (IOL) during the first 6 postoperative months.
SETTING
Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom.
DESIGN
Prospective case series.
METHOD: Eyes with age-related cataract and corneal astigmatism (1.00 to 3.00 D diopter [D]) were measured by partial coherence interferometry (IOLMaster). Preoperatively, the horizontal axis was marked. Surgical technique included a standardized temporal clear corneal incision, phacoemulsification, and in-the-bag implantation of a toric IOL (AcrySof SN60TT). Rotational stability of the IOL was assessed using retroillumination photographs postoperatively at 1 hour and at 1 and 6 months. Uncorrected (UDVA) and corrected distance visual acuities, residual refractive error, and keratometric and refractive cylinders were measured.
RESULTS
The study enrolled 30 eyes (30 patients). The mean absolute IOL rotation was 2.44 degrees ± 1.84 (SD) at 1 month and 2.66 ± 1.99 degrees at 6 months. The photographic technique had high reproducibility of axis measurement, with consecutive measurements varying by less than 2.0 degrees. The mean UDVA was 0.16 logMAR (range 0.42 to -0.18 logMAR) at 1 month and 0.20 logMAR (range 0.60 to -0.20 logMAR) at 6 months. The residual refractive cylinder by autorefraction was -0.84 D (range -2.00 to 0.00 D) and -0.80 D (range -1.75 to 0.00 D), respectively.
CONCLUSIONS
The acrylic toric IOL was rotationally stable within the first 6 months postoperatively. The photographic and axis analysis method to evaluate stability had high reproducibility and detected small changes in rotation.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.