DOI 10.1016/j.jcrs.2010.03.042

Медицинская библиотека · Journal of cataract and refractive surgery · 2010

Central corneal thickness measurements after myopic photorefractive keratectomy using Scheimpflug imaging, scanning-slit topography, and ultrasonic pachymetry.

Faramarzi Amir, Karimian Farid, Jafarinasab Mohammad Reza, Jabbarpoor Bonyadi Mohammad Hossein, Yaseri Mehdi

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare central corneal thickness (CCT) measurements obtained with a dual Scheimpflug camera system (Galilei), a scanning-slit topographer (Orbscan II), and an ultrasound (US) pachymeter (SP-2000) after photorefractive keratectomy (PRK) for myopia.

SETTING

Ophthalmic Research Center, Labbafinejad Medical Center, Tehran, Iran.

METHODS

This prospective study enrolled patients who had PRK to correct myopia or myopic astigmatism and a postoperative follow-up of at least 5 months. The CCT was measured in a single session using Scheimpflug imaging, scanning-slit topography, and US pachymetry. Data were analyzed with paired t tests, linear regression, and Bland-Altman plots.

RESULTS

The mean CCT in the 100 eyes (50 patients) was 524.06 microm +/- 38.56 (SD) with Scheimpflug imaging, 505.92 +/- 47.35 microm with scanning-slit topography, and 496.97 +/- 42.74 microm with US pachymetry. The lower to upper 95% limits of agreement (LoA) with US pachymetry were 2.4 to 51.8 microm for Scheimpflug imaging and -22.2 to 40.1 microm for scanning-slit topography. After application of a correction factor (0.95), the mean corrected Scheimpflug CCT measurement was 497.33 +/- 36.59 microm and the lower to upper 95% LoA with US pachymetry were -25.9 to 25.5 microm. The acoustic factor for scanning-slit topography was 0.94.

CONCLUSIONS

The CCT measurements in eyes that had PRK were thicker with Scheimpflug imaging than with US pachymetry or scanning-slit topography in the late postoperative period. With application of a correction factor, the Scheimpflug measurements were closer to US pachymetry values and had better agreement than scanning-slit topography.

Для цитирования: Faramarzi Amir, Karimian Farid, Jafarinasab Mohammad Reza, Jabbarpoor Bonyadi Mohammad Hossein, Yaseri Mehdi. Central corneal thickness measurements after myopic photorefractive keratectomy using Scheimpflug imaging, scanning-slit topography, and ultrasonic pachymetry.. Journal of cataract and refractive surgery. 2010. DOI: 10.1016/j.jcrs.2010.03.042.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.