DOI 10.1016/j.jcrs.2010.03.047

Медицинская библиотека · Journal of cataract and refractive surgery · 2010

Rotational and centration stability of an aspheric intraocular lens with a simulated toric design.

Buckhurst Phillip J, Wolffsohn James S, Naroo Shehzad A, Davies Leon N

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the stability of the Akreos AO intraocular lens (IOL) platform with a simulated toric design using objective image analysis.

SETTING

Six hospital eye clinics across Europe.

METHODS

After implantation in 1 eye of patients, IOLs with orientation marks were imaged at 1 to 2 days, 7 to 14 days, 30 to 60 days, and 120 to 180 days. The axis of rotation and IOL centration were objectively assessed using validated image analysis.

RESULTS

The study enrolled 107 patients with a mean age of 69.9 years +/- 7.7 (SD). The image quality was sufficient for IOL rotation analysis in 91% of eyes. The mean rotation between the first day postoperatively and 120 to 180 days was 1.93 +/- 2.33 degrees, with 96% of IOLs rotating fewer than 5 degrees and 99% rotating fewer than 10 degrees. There was no significant rotation between visits and no clear bias in the direction of rotation. In 71% of eyes, the dilation and image quality was sufficient for image analysis of centration. The mean change in centration between 1 day and 120 to 180 days was 0.21 +/- 0.11 mm, with all IOLs decentering less than 0.5 mm. There was no significant decentration between visits and no clear bias in the direction of the decentration.

CONCLUSION

Objective analysis of digital retroillumination images taken at different postoperative periods shows the aspheric IOL platform was stable in the eye and is therefore suitable for the application of a toric surface to correct corneal astigmatism.

Для цитирования: Buckhurst Phillip J, Wolffsohn James S, Naroo Shehzad A, Davies Leon N. Rotational and centration stability of an aspheric intraocular lens with a simulated toric design.. Journal of cataract and refractive surgery. 2010. DOI: 10.1016/j.jcrs.2010.03.047.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.