BACKGROUND
The activity of von Willebrand factor (VWF) in facilitating platelet adhesion and aggregation correlates with its multimer size. Traditional ristocetin-dependent functional assays lack sensitivity to multimer sizes. Recently, nanobodies targeting the autoinhibitory module and activating VWF were identified.
OBJECTIVES
To develop an assay that can differentiate the platelet-binding activity of VWF multimers.
METHODS
A novel enzyme-linked immunosorbent assay (nanobody-triggered glycoprotein Ib binding assay [VWF:GPIbNab]) utilizing a VWF-activating nanobody was developed. Recombinant VWF, plasma-derived VWF (pdVWF), and selected gel-filtrated fractions of pdVWF were evaluated for VWF antigen and activity levels. A linear regression model was developed to estimate the specific activity of VWF multimers.
RESULTS
Of the 3 activating nanobodies tested, 6C11 with the lowest activation effect exhibited the highest sensitivity for high-molecular-weight multimers (HMWMs) of VWF. VWF:GPIbNab utilizing 6C11 (VWF:GPIbNab6C11) produced significantly higher activity/antigen ratios for recombinant VWF (>2.0) and HMWM-enriched pdVWF fractions (>2.0) than for pdVWF (∼1.0) or fractions enriched with shorter multimers (<1.0). The differences were much larger than those produced by VWF:GPIbNab utilizing other nanobodies, VWF:GPIbM, VWF:GPIbR, or VWF:CB assays. Linear regression analysis of 5 pdVWF fractions of various multimer sizes produced an estimated specific activity of 2.7 for HMWMs. The analysis attributed >90% of the VWF activity measured by VWF:GPIbNab6C11 to that of HMWMs, which is significantly higher than all other activity assays tested.
CONCLUSION
The VWF:GPIbNab6C11 assay exhibits higher sensitivity to HMWMs than ristocetin-based and collagen-binding assays. Future studies examining the application of this assay in clinical settings and any associated therapeutic benefit of doing so are warranted.