Аннотация
PURPOSE OF REVIEW: Leadless pacemakers (LPs) are emerging as alternative cardiac implantable devices for the treatment of bradyarrhythmia. This article aims to review the data behind the safety and efficacy of these devices while highlighting their pros and cons.
RECENT FINDINGS: Prospective non-randomized studies and registries have found that LPs are associated with lower rate of device-related complications mainly driven by lower need for lead-related interventions as compared to traditional pacemakers. On the other hand, cardiac perforation appears to occur more frequently with LPs. LPs are associated with lower rate of device-related complications as compared to the traditional pacemakers. However, the rate of pericardial effusion is higher and is more severe. As we transition to multi-chamber LPs, it is important to ensure the safety and efficacy of these devices.