New-Generation Transcatheter Aortic Valves in Patients With Small Aortic Annuli - Comparison of Balloon- and Self-Expandable Valves in Asian Patients.

Okuyama Kazuaki, Izumo Masaki, Ochiai Tomoki, Kuwata Shingo, Kaihara Toshiki, Koga Masashi, Kamijima Ryo, Ishibashi Yuki, Tanabe Yasuhiro, Higuma Takumi, Makkar Raj, Miyairi Takeshi, Akashi Yoshihiro J

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Asian patients have smaller aortic annuli. Although 20-mm balloon-expandable (BE) transcatheter heart valves (THV) are manufactured for transcatheter aortic valve implantation (TAVI) in these cases, the supra-annular design of self-expandable (SE) THV is considered more suitable; however, real-world comparative data are scarce.

METHODS AND

RESULTS

Consecutive TAVI cases (n=330) in a single Japanese center were reviewed. Based on the cutoff for the new-generation 20-/23-mm BE-THV, a small aortic annulus was defined as <330 mm2. A considerable number of patients had small annuli: 49/302 (16%). Of these, 33 BE-THV and 13 SE-THV using new-generation valves were compared. Although the SE-THV group had smaller annulus area (median 297 (interquartile range, 280-313) vs. 309 (303-323) mm2(P=0.022)), it had more favorable post-procedural parameters; for SE-THV and BE-THV, respectively, effective orifice area (EOA), 1.5 (1.3-1.6) vs. 1.1 cm2(0.9-1.3) (P=0.002); mean pressure gradient, 7.6 (5.6-11.0) vs. 14.2 mmHg (11.2-18.8) (P=0.001); and peak velocity, 1.8 (1.6-2.4) vs. 2.7 m/s (2.3-3.1) (P=0.001). Although new left bundle branch block was higher with SE-THV (24% and 62%, P=0.02), patient-prosthesis mismatch (PPM) ≥ moderate (indexed EOA <0.85 cm2/m2) was significantly less with SE-THV than with BE-THV (8% vs. 55%; P=0.04). Hemodynamic findings were consistent up to 1 year.

CONCLUSIONS

Small annuli are often seen in Asian patients, for whom SE-THV implantation results in favorable hemodynamics with less PPM.

Для цитирования: Okuyama Kazuaki, Izumo Masaki, Ochiai Tomoki, Kuwata Shingo, Kaihara Toshiki, Koga Masashi, Kamijima Ryo, Ishibashi Yuki, Tanabe Yasuhiro, Higuma Takumi, Makkar Raj, Miyairi Takeshi, Akashi Yoshihiro J. New-Generation Transcatheter Aortic Valves in Patients With Small Aortic Annuli - Comparison of Balloon- and Self-Expandable Valves in Asian Patients.. Circulation journal : official journal of the Japanese Circulation Society. 2020. DOI: 10.1253/circj.cj-20-0368.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.