Journal of cardiovascular electrophysiology · 2020
Interrelationships between interventricular electrical delays in cardiac resynchronization therapy.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Moubarak Ghassan, Sebag Frédéric A, Socie Pierre, Villejoubert Olivier, Louembe Jules, Ferchaud Virginie
Аннотация
INTRODUCTION: In cardiac resynchronization therapy, pacing the left ventricle (LV) at sites of prolonged electrical delay is associated with better outcomes. We sought to characterize the interrelationships between intrinsic, right-ventricular (RV)-paced, and LV-paced interventricular delays.
METHODS AND RESULTS: The following electrical timings were measured at implantation for all electrodes of the LV quadripolar leads: QLV, interventricular delay in intrinsic rhythm (RVs-LVs), in RV-paced rhythm (RVp-LVs), and in LV-paced rhythm (LVp-RVs). We included 32 patients (78% men, age 72 years, LV ejection fraction 29%, left bundle branch block 84%). QLV and RVs-LVs were correlated (R2 = .72, p < .0001), as were RVs-LVs and RVp-LVs (R2 = .27, p = .002) and RVp-LVs and LVp-RVs (R2 = .60, p < .001). Direction of activation along the four LV lead electrodes was concordant between RVs-LVs and RVp-LVs in only 17 (53%) patients. The latest-activated electrodes in RVs-LVs and RVp-LVs were concordant in 26 (81%) patients, adjacent in 3 (9%) patients, and remote in 3 (9%) patients. Biventricular-paced QRS duration varied by more than 10 ms between the two electrodes in half of the patients with dissimilar latest electrodes. Among the seven echocardiographic nonresponders at 6 months, the programmed electrode was remote from the latest electrode in RVs-LVs in five patients and in RVp-LVs in three patients.
CONCLUSION: Intrinsic and RV-paced interventricular electrical delays are correlated, but there is substantial heterogeneity between patients. The latest-activated electrode may be different between RVs-LVs and RVp-LVs, and this might have important implications in selecting the optimal LV vector.
METHODS AND RESULTS: The following electrical timings were measured at implantation for all electrodes of the LV quadripolar leads: QLV, interventricular delay in intrinsic rhythm (RVs-LVs), in RV-paced rhythm (RVp-LVs), and in LV-paced rhythm (LVp-RVs). We included 32 patients (78% men, age 72 years, LV ejection fraction 29%, left bundle branch block 84%). QLV and RVs-LVs were correlated (R2 = .72, p < .0001), as were RVs-LVs and RVp-LVs (R2 = .27, p = .002) and RVp-LVs and LVp-RVs (R2 = .60, p < .001). Direction of activation along the four LV lead electrodes was concordant between RVs-LVs and RVp-LVs in only 17 (53%) patients. The latest-activated electrodes in RVs-LVs and RVp-LVs were concordant in 26 (81%) patients, adjacent in 3 (9%) patients, and remote in 3 (9%) patients. Biventricular-paced QRS duration varied by more than 10 ms between the two electrodes in half of the patients with dissimilar latest electrodes. Among the seven echocardiographic nonresponders at 6 months, the programmed electrode was remote from the latest electrode in RVs-LVs in five patients and in RVp-LVs in three patients.
CONCLUSION: Intrinsic and RV-paced interventricular electrical delays are correlated, but there is substantial heterogeneity between patients. The latest-activated electrode may be different between RVs-LVs and RVp-LVs, and this might have important implications in selecting the optimal LV vector.