His bundle pacing for newly acquired pacing needs in patients implanted with a subcutaneous implantable cardioverter defibrillator: A feasibility study based on the automated screening score and clinical cases.

Gauthey Anaïs, Calle Simon, Accinelli Stefano, Depuydt Philippe, Garnir Quentin, Scavée Christophe, Marchandise Sébastien, Wauters Aurélien, Bordachar Pierre, de Pooter Jan, le Polain de Waroux Jean-Benoît

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Management of subcutaneous implantable cardioverter defibrillator (S-ICD) patients with newly acquired pacing needs remains problematic. His bundle pacing (HBP) allows for cardiac pacing without significant changes in the QRS morphology. We hypothesized that HBP does not alter S-ICD sensing and functions.

METHODS

Twenty consecutive patients were implanted with a HB pacemaker. Among them, 17 demonstrated successful His recruitment and were prospectively screened with the automated screening tool (AST). Results of screenings performed immediately after implant and during follow-up, during intrinsic rhythm and while pacing from all available pacing configurations, were compared using the AST score. Positive-screening tests were defined by greater than or equal to 1 positive vector.

RESULTS

Among the 17 patients successfully implanted (male: 41%; mean age: 73), 13 presented an indication of ventricular pacing and four of cardiac resynchronization. Absolute AST scores during both HBP (all configurations) and intrinsic rhythm were similar (p: NS). Due to left bundle branch block correction, HBP resulted in higher number of positive vectors (AST ≥ 100). AST scores were higher during HBP when compared with right ventricular pacing (RVP) (primary vector: 272 [16; 648] vs 4.6 [0.8; 16.2]; P = .003; secondary vector: 569 [183; 1186] vs 1.5 [0.7; 8.3]; P < .0001; alternate vector: 44 [2;125] vs 4.8 [0.9; 9.3]; P = .02) and resulted in a much higher number of positive vectors. Up to 90% of the patients had a positive-screening test during HBP. This passing rate was higher when compared RVP (17%; P < .0001).

CONCLUSION

HBP restores normal intrinsic conduction and minimally modifies the surface electrocardiograph and subcutaneous electrograms. When ventricular pacing is needed, HBP might represent an ideal pacing option for patients implanted with a S-ICD.

Для цитирования: Gauthey Anaïs, Calle Simon, Accinelli Stefano, Depuydt Philippe, Garnir Quentin, Scavée Christophe, Marchandise Sébastien, Wauters Aurélien, Bordachar Pierre, de Pooter Jan, le Polain de Waroux Jean-Benoît. His bundle pacing for newly acquired pacing needs in patients implanted with a subcutaneous implantable cardioverter defibrillator: A feasibility study based on the automated screening score and clinical cases.. Journal of cardiovascular electrophysiology. 2020. DOI: 10.1111/jce.14566.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.