MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.athoracsur.2021.08.057

Медицинская библиотека · The Annals of thoracic surgery · 2021

Calcification of Gore-Tex Neochord After Mitral Repair: Electron Microscopy.

Luthra Suvitesh, Eissa Ahmed, Malvindi Pietro G, Tsang Geoffrey M

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

A 76-year-old woman presented with severe mitral valve regurgitation 6 years after mitral valve repair with Gore-Tex (expanded polytetrafluoroethylene, W.L. Gore & Associates Inc) neochords and ring annuloplasty. Echocardiography revealed a ruptured neochord. During successful mitral valve replacement, the explanted Gore-Tex neochords were found to be stiff and calcified, with a fracture. Electron microscopy was used to examine the explant and a control neochord. There was disruption of the microstructure with extensive calcium infiltration at the fracture point. Although this is a rare cause of late repair failure, it warrants yearly follow-up with echocardiography.

Для цитирования: Luthra Suvitesh, Eissa Ahmed, Malvindi Pietro G, Tsang Geoffrey M. Calcification of Gore-Tex Neochord After Mitral Repair: Electron Microscopy.. The Annals of thoracic surgery. 2021. DOI: 10.1016/j.athoracsur.2021.08.057.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.