Initial Experiences with the MitraClip G4: Review of the Novel Device Features.
Аннотация
OBJECTIVE
Percutaneous edge-to-edge mitral valve repair is an evolving procedure with increasing indications for use. The MitraClip™ G4 is the newest device with 4 device options, each with distinct characteristics. While the NTR and XTR clips have been in use, the G4 adds 2 new wider clips with additional features, the NTW and XTW.
METHODS
We reviewed 26 cases employing the MitraClip G4 at our institution from November 2019 to March 2020.
RESULTS
The NTR, XTR, NTW, and XTW clips were used in 3 (11.5%), 0 (0.0%), 16 (61.5%), and 11 (42.3%) cases, respectively. Median mitral regurgitation grade was reduced from 4+ to 1+, though 7 (27%) cases had 2+ residual mitral regurgitation.
CONCLUSIONS
The new G4 device includes simplified instructions for use, the option of a wider clip, controlled gripper actuation, and potential left atrial pressure monitoring. The newest MitraClip G4 device allows for potential improved treatment options for patients with mitral regurgitation.