Influence of edentulous span length on complete-arch intraoral scanner accuracy: a comparative in vitro study.

Mayer Yaniv, Berg Yarden, Kfir Or, Triestino Alessio, Gutmacher Zvi, Canullo Luigi

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

To assess the influence of edentulous span length on the accuracy of two intraoral scanner (IOS) devices Primescan (Dentsply Sirona) and Trios 3 (3Shape) in complete-arch scenarios. The hypothesis was that accuracy would decline with increasing span length, with performance differences between scanners.

METHOD AND MATERIALS: An in vitro study was conducted using a maxillary complete-arch model with a central incisor implant. Twelve span-length scenarios (FA1 to FA12) were created by sequentially removing adjacent teeth. Each condition was scanned 10 times using both IOS devices (n = 240). A laboratory scanner (Identica blue, Medit) provided reference scans. Trueness (vs reference) and precision (intra-group) were evaluated using root mean square (RMS) error. Scans were aligned via automatic and local best-fit in Medit Design. Statistical analysis included two-way mixed ANOVA, repeated measures ANOVA, and linear regression with Bonferroni correction (α = .004).

RESULTS

Trios 3 demonstrated lower overall RMS values (0.296 ± 0.043 mm) than Primescan (0.338 ± 0.049 mm), with significant differences in 9 of 12 scenarios (P .001). Correlation with span length was stronger for Primescan (R2 = 0.892) than Trios 3 (R2 = 0.674). Both devices showed high reliability.

CONCLUSIONS

Increasing edentulous span length negatively impacts scanner accuracy. Devices demonstrating consistent performance are preferable for reliable full-arch digital workflows. (Quintessence Int 2025;56:618-625; doi: 10.3290/j.qi.b639465).

Для цитирования: Mayer Yaniv, Berg Yarden, Kfir Or, Triestino Alessio, Gutmacher Zvi, Canullo Luigi. Influence of edentulous span length on complete-arch intraoral scanner accuracy: a comparative in vitro study.. Quintessence international (Berlin, Germany : 1985). 2025. DOI: 10.3290/j.qi.b639465.