The Annals of thoracic surgery · 2024

Hemodynamic Comparison of the On-X and Top Hat Mechanical Aortic Valve Prostheses.

СтатьяСравнительное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Juarez-Casso Fernando M, Cangut Busra, King Katherine S, Lee Alex T, Stulak John M, Schaff Hartzell V, Greason Kevin L

Аннотация

BACKGROUND: There are limited data comparing hemodynamic valve function in mechanical aortic valve prostheses. This study compared the hemodynamic function of 2 commonly used mechanical aortic valve (AV) prostheses, the On-X (Artivion) and Top Hat (CarboMedics Inc) valves.

METHODS: This study was a retrospective analysis of 512 patients who underwent AV replacement with the On-X (n = 252; 49%) or Top Hat (n = 260; 51%) mechanical valves between 2011 and 2019. Patients were matched on the basis of selected variables. Echocardiographic data were collected preoperatively and postoperatively over a median follow-up of 1.39 years.

RESULTS: A total of 320 patients were matched, 160 patients in each group. Despite being matched for left ventricular outflow tract diameter, patients in the Top Hat group received a greater prevalence of smaller tissue annulus diameter valves (≤21 mm) (83% vs 38%; P < .001). Patients in the On-X group had longer aortic cross-clamp times (78 minutes vs 64 minutes; P < .001) during isolated aortic valve replacement. Discharge echocardiography showed no difference in the AV area index between both groups (1.00 cm2/m2 vs 1.02 cm2/m2; P = .377). During longer-term echocardiographic follow-up, the AV area index remained stable for both valves within their respective tissue annulus diameter groups (P = .060).

CONCLUSIONS: There was no difference between the 2 valves with respect to the AV area index at discharge, and hemodynamic function was stable during longer-term follow-up. The longer aortic cross-clamp time observed in the On-X group may indicate increased complexity of implantation compared with the Top Hat group.