OBJECTIVES

Severe aortic stenosis (AS) in bicuspid aortic valves (BAV) is associated with an increased risk of paravalvular leak (PVL) after a transcatheter aortic valve replacement (TAVR). Virtual reality (VR) has been shown to be an effective tool in surgical training, but its utility in clinical practice has not been studied. Here we present the first study to evaluate the use of VR simulation in pre-procedure planning and prediction of PVL in TAVR in patients with severe BAV AS.

METHODS

Twenty-two patients with severe BAV AS undergoing TAVR between 2014 and 2018 at the University of Minnesota were included in the study. VR simulation of TAVR implants was performed and implants were analyzed for PVL. The primary endpoint was the percent circumference of valve malapposition in VR as compared to the severity of PVL on post-procedure echocardiography.

RESULTS

The median age was 78.26 years (IQR 63.77-86.79) and 40.9% (n = 9) were female. Our VR model accurately predicted the presence and absence of PVL in all patients (17/17 and 5/5, respectively). The mean circumferential PVL was 3.73 % ± 7.71. The receiver operator characteristic curve showed an area under the curve of 0.83 (0.59-1.00, P = .03) for malapposition in the VR-TAVR simulated model.

CONCLUSIONS

VR-TAVR implantation may predict PVL in severe BAV AS undergoing TAVR.

Для цитирования: Chahine Johnny, Mascarenhas Lorraine, Yannopoulos Demetris, Raveendran Ganesh, Gurevich Sergey. Virtual reality to predict paravalvular leak in bicuspid severe aortic valve stenosis in transcatheter aortic valve implants.. The Journal of invasive cardiology. 2024. DOI: 10.25270/jic/24.00019.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.