DOI 10.20471/acc.2024.63.s1.16
VALVE-IN-VALVE IMPLANTATION GONE SOUTH: A CASE REPORT.
Аннотация
Valve-in-Valve (ViV) procedures for failed surgical bioprostheses are becoming increasingly common. We present a case of a 73-year-old male patient with significant comorbidities who underwent a primary transcatheter aortic valve implantation (TAVI) with an Evolut 29 mm (Medtronic, Minneapolis, MN, USA) valve. This TAVI-prosthesis implantation failed due to the inability to fix the valve at the appropriate height to achieve an optimal hemodynamic result. The valve had to be snared and pulled into an adequate position and was then fixed by an implantation of another TAVI-prosthesis (29 mm Sapien S3, Edwards Lifesciences, Irvine, CA, USA) with optimal final result. The patient was discharged after significant clinical improvement and without further complications.