DOI 10.1016/j.ijid.2020.06.076
Comparison of diagnostic accuracy of the GeneXpert Ultra and cell-free nucleic acid assay for tuberculous meningitis: A multicentre prospective study.
Аннотация
DESIGN
A prospective multicentre study was conducted to compare the diagnostic accuracy of the GeneXpert MTB/RIF Ultra (Xpert Ultra) and cell-free DNA (cfDNA) assay for tuberculous meningitis (TBM) in China.
METHODS
A prospective cohort study was conducted among individuals with symptoms suggestive of TBM registered in three TB specialised hospitals in China between June 2018 and January 2019.
RESULTS
Overall, 84 patients suggestive of TBM were included in this analysis between June 2018 and January 2019. Using microbiological evidence as reference, the sensitivity/specificity for the diagnostic tests were Xpert Ultra 93.3%/100%, cfDNA 93.3%/92.6% and mycobacteria growth indicator tube (MGIT) culture 13.3%/100%. In addition, the sensitivity of culture was 6.7% when using clinical diagnosis criteria as the gold standard. Xpert Ultra correctly identified 28 cases as TBM, indicating a sensitivity of 46.7%. Notably, four additional TBM cases were detected by cfDNA compared with Xpert Ultra, yielding an overall sensitivity of 53.3%. Statistical analysis revealed that the sensitivity of Xpert Ultra and cfDNA was significantly higher than that of culture.
CONCLUSIONS
The data demonstrate that Xpert Ultra and cfDNA assay showed optimal sensitivity compared with MGIT culture. In addition, there was no significant correlation between bacterial load and TBM severity in the participants.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.