Application of a novel endoscopic ultrasound-guided motorized fine needle for liver and pancreatic lesion biopsy.
Аннотация
BACKGROUND AND AIMS: The motorized fine-needle biopsy (mFNB) device (Precision-GI; Limaca Medical, Ein Ha Emeq, Israel) is a novel electromechanical tool designed to optimize EUS-guided tissue acquisition through controlled axial advancement and high-speed rotational coring. This study presents 2 clinical cases demonstrating the safety and effectiveness of the second-generation mFNB in obtaining liver and pancreatic biopsy samples.
METHODS
Two female patients underwent EUS-guided biopsies using a 20-gauge mFNB device. One patient had suspected liver disease and the other a pancreatic mass, respectively. The device was used to perform targeted tissue sampling, with emphasis on tissue adequacy, integrity, and absence of blood contamination.
RESULTS
In both cases, the mFNB allowed successful tissue acquisition without adverse events. The liver biopsy yielded a 0.5 × 0.5 × 0.2-cm specimen with 6 portal triads, confirming micronodular cirrhosis with steatosis and fibrosis. The pancreatic biopsy produced a 0.8 × 0.5 × 0.2-cm core, leading to a diagnosis of acinar cell carcinoma. Both specimens were deemed adequate and intact with no blood contamination.
CONCLUSIONS
The mFNB device is safe for EUS-guided liver and pancreatic biopsies, providing high-quality tissue samples. The motorized design may reduce procedure time and improve diagnostic yield. Further comparative studies with manual fine-needle biopsy are required to validate these findings.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.