Journal of magnetic resonance imaging : JMRI · 2024

Reproducibility and Repeatability of Intravoxel Incoherent Motion MRI Acquisition Methods in Liver.

СтатьяПоддержка исследования: вне правительства СШАПоддержка NIH: внешняя программаНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Vasquez Juan A, Brown Marissa, Woolsey Mary, Abdul-Ghani Mohammad, Katabathina Venkata, Deng Shengwen, Blangero John, Clarke Geoffrey D

Аннотация

BACKGROUND: Intravoxel incoherent motion (IVIM) diffusion weighted MRI (DWI) has potential for evaluating hepatic fibrosis but image acquisition technique influence on diffusion parameter estimation bears investigation.

PURPOSE: To minimize variability and maximize repeatably in abdominal DWI in terms of IVIM parameter estimates.

STUDY TYPE: Prospective test-retest and image quality comparison.

SUBJECTS: Healthy volunteers (3F/7M, 29.9 ± 12.9 years) and Family Study subjects (18F/12M, 51.7 ± 16.7 years), without and with liver steatosis.

FIELD STRENGTH/SEQUENCE: Abdominal single-shot echo-planar imaging (EPI) and simultaneous multi-slice (SMS) DWI sequences with respiratory triggering (RT), breath-holding (BH), and navigator echo (NE) at 3 Tesla.

ASSESSMENT: SMS-BH, EPI-NE, and SMS-RT data from twice-scanned healthy volunteers were analyzed using 6 × b-values (0-800 s⋅mm-2) and lower (LO) and higher (HI) b-value ranges. Family Study subjects were scanned using SMS and standard EPI sequences. The biexponential IVIM model was used to estimate fast-diffusion coefficient (Df), fraction of fast diffusion (f), and slow-diffusion coefficient (Ds). Scan time, estimated signal-to-noise ratio (eSNR), eSNR per acquisition, and distortion ratio were compared.

STATISTICAL TESTS: Coefficients of variation (CoV) and Bland Altman analyses were performed for test-retest repeatability. Interclass correlation coefficient (ICC) assessed interobserver agreement with P < 0.05 deemed significant.

RESULTS: Within-subject CoVs among volunteers (N = 10) for f and Ds were lowest in EPI-NE-LO (11.6%) and SMS-RT-HI (11.1%). Inter-observer ICCs for f and Ds were highest for EPI-NE-LO (0.63) and SMS-RT-LO (0.76). Df could not be estimated for most subjects. Estimated eSNR (EPI = 21.9, SMS = 4.7) and eSNR time (EPI = 6.7, SMS = 16.6) were greater for SMS, with less distortion in the liver region (DR-PE: EPI = 23.6, SMS = 13.1).

DATA CONCLUSION: Simultaneous multislice acquisitions had significantly less variability and higher ICCs of Ds, higher eSNR, less distortion, and reduced scan time compared to EPI.

EVIDENCE LEVEL: 2 TECHNICAL EFFICACY: Stage 1.