DOI 10.1002/jmri.29249

Медицинская библиотека · Journal of magnetic resonance imaging : JMRI · 2024

Reproducibility and Repeatability of Intravoxel Incoherent Motion MRI Acquisition Methods in Liver.

Vasquez Juan A, Brown Marissa, Woolsey Mary, Abdul-Ghani Mohammad, Katabathina Venkata, Deng Shengwen, Blangero John, Clarke Geoffrey D

СтатьяПоддержка исследования: вне правительства СШАПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

BACKGROUND

Intravoxel incoherent motion (IVIM) diffusion weighted MRI (DWI) has potential for evaluating hepatic fibrosis but image acquisition technique influence on diffusion parameter estimation bears investigation.

PURPOSE: To minimize variability and maximize repeatably in abdominal DWI in terms of IVIM parameter estimates.

STUDY TYPE: Prospective test-retest and image quality comparison.

SUBJECTS: Healthy volunteers (3F/7M, 29.9 ± 12.9 years) and Family Study subjects (18F/12M, 51.7 ± 16.7 years), without and with liver steatosis.

FIELD STRENGTH/SEQUENCE: Abdominal single-shot echo-planar imaging (EPI) and simultaneous multi-slice (SMS) DWI sequences with respiratory triggering (RT), breath-holding (BH), and navigator echo (NE) at 3 Tesla.

ASSESSMENT: SMS-BH, EPI-NE, and SMS-RT data from twice-scanned healthy volunteers were analyzed using 6 × b-values (0-800 s⋅mm-2) and lower (LO) and higher (HI) b-value ranges. Family Study subjects were scanned using SMS and standard EPI sequences. The biexponential IVIM model was used to estimate fast-diffusion coefficient (Df), fraction of fast diffusion (f), and slow-diffusion coefficient (Ds). Scan time, estimated signal-to-noise ratio (eSNR), eSNR per acquisition, and distortion ratio were compared.

STATISTICAL TESTS: Coefficients of variation (CoV) and Bland Altman analyses were performed for test-retest repeatability. Interclass correlation coefficient (ICC) assessed interobserver agreement with P < 0.05 deemed significant.

RESULTS

Within-subject CoVs among volunteers (N = 10) for f and Ds were lowest in EPI-NE-LO (11.6%) and SMS-RT-HI (11.1%). Inter-observer ICCs for f and Ds were highest for EPI-NE-LO (0.63) and SMS-RT-LO (0.76). Df could not be estimated for most subjects. Estimated eSNR (EPI = 21.9, SMS = 4.7) and eSNR time (EPI = 6.7, SMS = 16.6) were greater for SMS, with less distortion in the liver region (DR-PE: EPI = 23.6, SMS = 13.1).

DATA

CONCLUSION

Simultaneous multislice acquisitions had significantly less variability and higher ICCs of Ds, higher eSNR, less distortion, and reduced scan time compared to EPI.

EVIDENCE LEVEL: 2 TECHNICAL EFFICACY: Stage 1.

Для цитирования: Vasquez Juan A, Brown Marissa, Woolsey Mary, Abdul-Ghani Mohammad, Katabathina Venkata, Deng Shengwen, Blangero John, Clarke Geoffrey D. Reproducibility and Repeatability of Intravoxel Incoherent Motion MRI Acquisition Methods in Liver.. Journal of magnetic resonance imaging : JMRI. 2024. DOI: 10.1002/jmri.29249.