MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/ijg.0000000000002023

Медицинская библиотека · Journal of glaucoma · 2022

Evaluating Intraocular Pressure After Myopic Photorefractive Keratectomy: A Comparison of Different Tonometers.

Lanza Michele, Sbordone Sandro, Tortori Antonia, Gironi Carnevale Ugo A, Melillo Paolo, Simonelli Francesca

СтатьяПеревод пока не подготовлен

Аннотация

PRCIS: All devices evaluated in this study showed a significant underestimation in intraocular pressure (IOP) measurement after myopic photorefractive keratectomy (PRK), Goldmann tonometer more than others. Corneal biomechanics indicated a different influence on the different kinds of tonometry.

PURPOSE: The aim was to investigate the reliability of Goldmann applanation tonometry (GAT), rebound tonometry (RT), ocular response analyzer (ORA), Corvis ST (CST), and dynamic contour tonometry (DCT) in IOP evaluation after myopic PRK.

MATERIALS AND

METHODS

One eye of 145 patients who underwent myopic PRK for a refractive defect ranging from -10.25 to -0.50 D (mean -4.69±2.00 D) was included in this retrospective comparative study. A complete eye visit with corneal tomography and IOP measurement with GAT, DCT, ORA, RT, and CST was performed before surgery and at 1, 3, and 6 months follow-up. Values provided by each device were tested and compared at each follow-up. Correlation analyses were run between changes in IOP and the corneal, morphologic and biomechanical parameters were measured after PRK.

RESULTS

GAT, DCT, ORA, RT, and CST showed a significant (P<0.01) underestimation of IOP at 6 months follow-up. GAT showed the greatest underestimation (-14.1%) and stronger correlations with corneal deformation parameter changes, whereas ORA, DCT, and RT appeared to be less conditioned by these variations. At 6 months follow-up DCT, ORA, RT, and CST provided IOP values with nonsignificant differences compared with GAT before PRK.

CONCLUSIONS

Each tested tonometer showed a significant IOP underestimation after myopic PRK. As this was most observed with GAT compared with all devices, we suggest DCT, ORA, RT, or CST to evaluate IOP in these patients following surgery.

Для цитирования: Lanza Michele, Sbordone Sandro, Tortori Antonia, Gironi Carnevale Ugo A, Melillo Paolo, Simonelli Francesca. Evaluating Intraocular Pressure After Myopic Photorefractive Keratectomy: A Comparison of Different Tonometers.. Journal of glaucoma. 2022. DOI: 10.1097/ijg.0000000000002023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.