MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/j.jcrs.0000000000000950

Медицинская библиотека · Journal of cataract and refractive surgery · 2022

Factors leading to realignment or exchange after implantable collamer lens implantation in 10 258 eyes.

Wei Ruoyan, Li Meiyan, Aruma Aruma, Knorz Michael C, Yang Dong, Yu Yongfu, Wang Xiaoying, Choi Joanne, Yao Peijun, Zhou Xingtao

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the incidence, causes, and outcomes of implantable collamer lens (ICL) realignment or exchange after implantation of the EVO-ICL.

SETTING

Eye and ENT Hospital of Fudan University, China.

DESIGN

Retrospective study.

METHODS

The study included 10 258 consecutive eyes after ICL implantation. Preoperative refractive and biometric measurements were compared between eyes with and without realignment or exchange. For eyes with ICL realignment or exchange, visual and biometric outcomes were also compared before and after ICL realignment or exchange.

RESULTS

The overall incidence of ICL realignment or exchange was 0.21% 22 eyes. 12 eyes (0.12%) underwent ICL realignment by axis rotation (10 eyes) or ICL exchange (2 eyes) due to toric ICL (TICL) misalignment. After realignment, uncorrected distance visual acuity improved, and residual cylinder decreased from -1.75 ± 0.48 diopters (D) to -0.87 ± 0.59 D ( P = .01). 10 eyes (0.10%) underwent vertical rotation of ICL (3 eyes) or ICL exchange (7 eyes) due to excessive vault. After either vertical rotation or ICL exchange, vault decreased significantly ( P < .05).

CONCLUSIONS

The incidence of ICL realignment or exchange after ICL implantation is low. TICL misalignment and excessive vault are 2 main causes. Implant exchange may be performed for excessive vault or misalignment with an insufficient vault. In addition, vertical rotation of an ICL may be a less invasive method to treat excessive vault in certain cases.

Для цитирования: Wei Ruoyan, Li Meiyan, Aruma Aruma, Knorz Michael C, Yang Dong, Yu Yongfu, Wang Xiaoying, Choi Joanne, Yao Peijun, Zhou Xingtao. Factors leading to realignment or exchange after implantable collamer lens implantation in 10 258 eyes.. Journal of cataract and refractive surgery. 2022. DOI: 10.1097/j.jcrs.0000000000000950.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.