Clinical outcome and paravalvular leakage of the new balloon-expandable Edwards Sapien 3 valve in comparison to its predecessor model (Edwards Sapien XT) in patients undergoing transfemoral aortic valve replacement.

Bocksch Wolfgang, Grossmann Benita, Geisler Tobias, Steeg Martin, Droppa Michal, Jorbenadze Rezo, Haap Michael, Gawaz Meinrad, Fateh-Moghadam Suzanne

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

The aim of this study was to compare the 30-day procedural, clinical and echocardiographic outcome of the new balloon-expandable Edwards Sapien 3 (ES3) valve with the Edwards Sapien XT (ESXT).

BACKGROUND

Post-implant paravalvular leaks (PVL) after transfemoral aortic valve replacement (TAVR) resulting in residual aortic regurgitation (AR) are a major limitation for long term outcome. New TAVR-devices have to eliminate this problem.

METHODS

Transfemoral TAVR was performed in 209 consecutive intermediate-high-risk surgical patients (pts) with symptomatic aortic stenosis (ESXT n = 102, ES3 n = 107). Transthoracic echocardiography (TTE) and 3-dimensional computed tomography were used for valve size selection. Primary endpoint of the study was none/trace AR derived by TTE 30-days after TAVR.

RESULTS

All pts underwent successfully TAVR with a combined device success of 100/102 (99%) in ESXT and 107/107 (100%) in ES3 pts. Fluoroscopy time (ESXT 11.8 ± 0.5 min vs. ES3 10.0 ± 0.5 min, P = 0.003) and contrast (ESXT 188.9 ± 5.6 mL vs. ES3 170.4 ± 4.7 mL, P = 0.04) were significantly lower in ES3 patients. 30-day clinical events did not differ. Transvalvular mean pressure gradients were significantly reduced to 7.4 ± 0.8 mmHg after ESXT and to 10.1± 0.4 mmHg after ES3 implantation. After 30 days none/trace AR was found in 34.3% (n = 35) of all ESXT pts in contrast to 89.7% (n = 96) of all ES3 patients. Moderate-to-severe AR was found rarely (ESXT 2.9% vs. ES3 0%, P = 0.073).

CONCLUSIONS

Although there was no significant difference in 30 day mortality, the newer ES3 valve reduced significantly residual paravalvular leakage. © 2016 Wiley Periodicals, Inc.

Для цитирования: Bocksch Wolfgang, Grossmann Benita, Geisler Tobias, Steeg Martin, Droppa Michal, Jorbenadze Rezo, Haap Michael, Gawaz Meinrad, Fateh-Moghadam Suzanne. Clinical outcome and paravalvular leakage of the new balloon-expandable Edwards Sapien 3 valve in comparison to its predecessor model (Edwards Sapien XT) in patients undergoing transfemoral aortic valve replacement.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2016. DOI: 10.1002/ccd.26562.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.