Surface ablation with iris recognition and dynamic rotational eye tracking-based tissue saving treatment with the Technolas 217z excimer laser.

Prakash Gaurav, Agarwal Amar, Kumar Dhivya Ashok, Jacob Soosan, Agarwal Athiya, Maity Amrita

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the visual and refractive outcomes and expected benefits of Tissue Saving Treatment algorithm-guided surface ablation with iris recognition and dynamic rotational eye tracking.

METHODS

This prospective, interventional case series comprised 122 eyes (70 patients). Pre- and postoperative assessment included uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), refraction, and higher order aberrations. All patients underwent Tissue Saving Treatment algorithm-guided surface ablation with iris recognition and dynamic rotational eye tracking using the Technolas 217z 100-Hz excimer platform (Technolas Perfect Vision GmbH). Follow-up was performed up to 6 months postoperatively. Theoretical benefit analysis was performed to evaluate the algorithm's outcomes compared to others.

RESULTS

Preoperative spherocylindrical power was sphere -3.62 ± 1.60 diopters (D) (range: 0 to -6.75 D), cylinder -1.15 ± 1.00 D (range: 0 to -3.50 D), and spherical equivalent -4.19 ± 1.60 D (range: -7.75 to -2.00 D). At 6 months, 91% (111/122) of eyes were within ± 0.50 D of attempted correction. Postoperative UDVA was comparable to preoperative CDVA at 1 month (P=.47) and progressively improved at 6 months (P=.004). Two eyes lost one line of CDVA at 6 months. Theoretical benefit analysis revealed that of 101 eyes with astigmatism, 29 would have had cyclotorsion-induced astigmatism of ≥ 10% if iris recognition and dynamic rotational eye tracking were not used. Furthermore, the mean percentage decrease in maximum depth of ablation by using the Tissue Saving Treatment was 11.8 ± 2.9% over Aspheric, 17.8 ± 6.2% over Personalized, and 18.2 ± 2.8% over Planoscan algorithms.

CONCLUSIONS

Tissue saving surface ablation with iris recognition and dynamic rotational eye tracking was safe and effective in this series of eyes.

Для цитирования: Prakash Gaurav, Agarwal Amar, Kumar Dhivya Ashok, Jacob Soosan, Agarwal Athiya, Maity Amrita. Surface ablation with iris recognition and dynamic rotational eye tracking-based tissue saving treatment with the Technolas 217z excimer laser.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2010. DOI: 10.3928/1081597x-20100428-01.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.