Journal of refractive surgery (Thorofare, N.J. : 1995) · 2010
LASIK for myopia and astigmatism using the SCHWIND AMARIS excimer laser: an international multicenter trial.
Контролируемое клиническое исследованиеСтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Arbelaez Maria Clara, Aslanides Ioannis M, Barraquer Carmen, Carones Francesco, Feuermannova Alena, Neuhann Tobias, Rozsival Pavel
Аннотация
PURPOSE: To assess the efficacy, predictability, and safety of LASIK for the surgical correction of low to moderate myopia with astigmatism using the SCHWIND AMARIS excimer laser.
METHODS: Six international study sites enrolled 358 eyes with a manifest refraction spherical equivalent (MRSE) from -0.50 to -7.38 diopters (D) (mean sphere: -3.13+/-1.58 D) with up to -5.00 D of astigmatism (mean: -0.69+/-0.67 D). All eyes underwent treatment with the nonwavefront-guided aspheric algorithm of the SCHWIND AMARIS excimer laser. All eyes were targeted for emmetropia. Refractive outcomes and corneal higher order aberrations were analyzed pre- and postoperatively. Visual quality was assessed using photopic and mesopic contrast sensitivity. Six-month postoperative outcomes are reported.
RESULTS: At 6 months postoperative, the MRSE for all eyes was -0.21+/-0.20 D, and 96% (343/358) of eyes had MRSE within +/-0.50 D. Uncorrected visual acuity was 20/20 or better in 98% (351/358) of eyes, and no eyes lost 2 or more lines of best spectacle-corrected visual acuity. The total corneal higher order aberrations root-mean-square increased by 0.09 microm, spherical aberration increased by 0.08 microm, and coma increased by 0.04 microm postoperatively. Photopic and mesopic contrast sensitivity did not change 6 months postoperatively.
CONCLUSIONS: Treatment of myopia with astigmatism using the SCHWIND AMARIS excimer laser is safe, efficacious, predictable, and maintains visual quality.
METHODS: Six international study sites enrolled 358 eyes with a manifest refraction spherical equivalent (MRSE) from -0.50 to -7.38 diopters (D) (mean sphere: -3.13+/-1.58 D) with up to -5.00 D of astigmatism (mean: -0.69+/-0.67 D). All eyes underwent treatment with the nonwavefront-guided aspheric algorithm of the SCHWIND AMARIS excimer laser. All eyes were targeted for emmetropia. Refractive outcomes and corneal higher order aberrations were analyzed pre- and postoperatively. Visual quality was assessed using photopic and mesopic contrast sensitivity. Six-month postoperative outcomes are reported.
RESULTS: At 6 months postoperative, the MRSE for all eyes was -0.21+/-0.20 D, and 96% (343/358) of eyes had MRSE within +/-0.50 D. Uncorrected visual acuity was 20/20 or better in 98% (351/358) of eyes, and no eyes lost 2 or more lines of best spectacle-corrected visual acuity. The total corneal higher order aberrations root-mean-square increased by 0.09 microm, spherical aberration increased by 0.08 microm, and coma increased by 0.04 microm postoperatively. Photopic and mesopic contrast sensitivity did not change 6 months postoperatively.
CONCLUSIONS: Treatment of myopia with astigmatism using the SCHWIND AMARIS excimer laser is safe, efficacious, predictable, and maintains visual quality.