BACKGROUND

Recent advancements in implant technology offer updated options for surgical management that have been rapidly adopted into clinical practice. The objective of this study is to biomechanically test and compare the current fixation options available for surgical fixation of two-part proximal humerus fractures and establish load to failure and stiffness values.

METHODS

Sixteen match-paired (32 total) fresh-frozen, cadaveric specimens were randomized to receive 1 of 4 fixation constructs following creation of an AO/OTA Type 11A3 (two-part) proximal humerus fractures. Fixation constructs tested consisted of 3.5 mm fixed angle plate (3.5-FAP), 4.5 mm fixed angle plate (4.5-FAP), humeral intramedullary nail (IMN), and a humeral intramedullary nail with a fixed angle blade (IMN-FAB). Specimen bone density was measured to ensure no adequate, non-osteoporotic bone. Constructs were tested for stiffness and ultimate load to failure and compared via one-way ANOVA analysis with subsequent post hoc Tukey HSD multiple group comparison statistical analysis.

RESULTS

The IMN-FAB construct was significantly stiffer than the 3.5-FAP construct (123.8 vs. 23.9, p<0.0001), the 4.5-FAP construct (123.8 vs. 33.3, p<0.0001) and the IMN construct (123.8 vs. 60.1, p=0.005). The IMN-FAB construct reported a significantly higher load to failure than the 3.5-FAB construct (4667.3 N vs. 1756.9 N, p<0.0001), and the 4.5-FAP construct (4667.3 N vs. 2829.4 N, p=0.019, Table 2). The IMN construct had a significantly higher load to failure than the 3.5-FAP construct (3946.8 vs. 1756.9, p=0.001, Table 2).

CONCLUSION

Biomechanical testing of modern fixation options for two-part proximal humerus fracture exhibited that the stiffest and highest load to failure construct was the IMN-FAB followed by the IMN, 3.5-FAP and then the 4.5-FAP constructs. However, prospective clinical trials with longer-term follow-up are required for definitive assessment of the ideal fixation construct for surgical management of two-part proximal humerus fractures.

Для цитирования: Yoon Richard S, Dziadosz Daniel, Porter David A, Frank Matthew A, Smith Wade R, Liporace Frank A. A comprehensive update on current fixation options for two-part proximal humerus fractures: a biomechanical investigation.. Injury. 2013. DOI: 10.1016/j.injury.2013.08.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.