MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/23259671241279847

Медицинская библиотека · Orthopaedic journal of sports medicine · 2024

Biomechanical Comparison of Suture-Relay All-Suture Anchors and Conventional Suture Anchors for Posterior Medial Meniscus Root Repair in Porcine Models.

Saengpetch Nadhaporn, Prasitmeeboon Napat, Janyawongchot Tanapol, Aroonjarattham Panya, Somtua Chompunut, Thamyongkit Sorawut

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Posterior medial meniscus root (PMMR) tears (PMMRTs) can be repaired using various techniques to promote healing. However, the biomechanical properties of suture-relay all-suture anchor (ASA) versus conventional suture anchor (CSA) and loop-locking transtibial pullout (TTP) have not been well established.

PURPOSE: To compare the biomechanical properties of PMMR repairs using suture-relay ASA, CSA, and loop-locking TTP.

STUDY

DESIGN

Controlled laboratory study.

METHODS

A total of 33 fresh-frozen porcine knee joints with intact medial menisci were randomly divided into 3 groups, with 11 specimens in each group: ASA, CSA, and TTP. The study involved cyclic loading, with displacement measurements taken after 100, 500, and 1000 cycles. Subsequently, the specimens were loaded until clinical failure (defined as 3-mm displacement) and then to ultimate failure of the construct, with data recorded for displacement after cyclic loading, load to 3-mm displacement, and ultimate load to failure.

RESULTS

After 1000 cyclic loadings, the suture-relay ASA group showed considerably less displacement than the loop-locking TTP group (1.8 ± 0.7 mm vs 2.9 ± 0.3 mm; P < .001), but the displacements did not differ considerably between the suture-relay ASA and CSA groups (2.2 ± 0.9 mm; P > .05). The mean loads to clinical failure were significantly greater in the suture-relay ASA and CSA groups (61.3 ± 6.5 and 57.5 ± 9.7 N, respectively) than in the loop-locking TTP group (38.3 ± 9.4 N; P < .05). The ultimate load to failure was significantly greater in the suture-relay ASA group than in the loop-locking TTP group (153 ± 55.1 N vs 102 ± 12.9 N; P < .05). All specimens in the loop-locking TTP group failed by suture elongation mode, whereas only 2 specimens (18%) in the suture-relay ASA group and 5 specimens (45%) in the CSA group failed by suture elongation. Nine specimens (82%) in the suture-relay ASA group and 6 specimens (55%) in the CSA group failed due to suture cutout through the meniscal tissue.

CONCLUSION

The biomechanical properties after PMMR repair did not statistically differ between the suture-relay ASA and CSA groups. The suture-relay ASA technique had a higher load to failure than the loop-locking TTP technique.

CLINICAL RELEVANCE: The suture-relay ASA technique is a promising option for the repair of PMMRTs; its repairing strength is also comparable to that of the CSA technique. Notably, the suture-relay ASA technique can be utilized without establishing a posteromedial portal, resulting in decreased procedure time and mitigating challenges associated with working from the posterior aspect of the knee.

Для цитирования: Saengpetch Nadhaporn, Prasitmeeboon Napat, Janyawongchot Tanapol, Aroonjarattham Panya, Somtua Chompunut, Thamyongkit Sorawut. Biomechanical Comparison of Suture-Relay All-Suture Anchors and Conventional Suture Anchors for Posterior Medial Meniscus Root Repair in Porcine Models.. Orthopaedic journal of sports medicine. 2024. DOI: 10.1177/23259671241279847.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.