MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ajo.2023.06.003

Медицинская библиотека · American journal of ophthalmology · 2023

Objective Quantification of Lens Opacity in Posterior Subcapsular Cataracts Using IOL Master 700 and CASIA-2.

Fan Shuxin, Feng Rui, Liang Feiyan, Li Chuang, Tao Shuya, Wang Mingwei, He Chang, Liu Xialin

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To objectively quantify the lens opacity of posterior subcapsular cataracts (PSCs) using the swept-source optical coherence tomography (SS-OCT)-based devices including IOL Master 700 and CASIA-2.

DESIGN

Prospective cross-sectional study.

METHODS

A total of 101 eyes of 101 patients with PSCs were enrolled in Zhongshan Ophthalmic Center from 2021 to 2022. The IOL Master 700 and CASIA-2 were used to obtain lens images. The average posterior subcapsular density (APSD) and the maximum posterior subcapsular density (MPSD) within the pupil area (radius: 3 or 5 mm) were measured by Image J. Spearman and Pearson correlation analysis were performed to assess the associations.

RESULTS

APSD-3mm, APSD-5mm, MPSD-3mm, and MPSD-5mm had positive correlations with best corrected visual acuity (BCVA; r = 0.658, 0.641, 0.583, and 0.572, P < .001, respectively), all of which were higher than the correlation between LOCS-III P score and BCVA (r = 0.548, P < .001). Particularly, the APSD-3mm showed the highest correlation with BCVA. APSD could distinguish severe PSCs (LOCS-III P score ≥ 5) with an area under the receiver operating characteristic curve (AUC) of 0.836 (95% CI 0.743-0.930) for APSD-3mm and with AUC 0.758 (95% CI 0.643-0.873) for APSD-5mm, highlighting the better performance of APSD-3mm. The APSD-3mm of IOL Master 700 correlated strongly with that of CASIA-2 (r = 0.789, P < .001).

CONCLUSIONS

This study presented an objective method for quantifying PSCs using IOL Master 700 and CASIA-2. APSD-3mm can be used as a new accurate and objective index for the quantitative assessment of PSCs.

Для цитирования: Fan Shuxin, Feng Rui, Liang Feiyan, Li Chuang, Tao Shuya, Wang Mingwei, He Chang, Liu Xialin. Objective Quantification of Lens Opacity in Posterior Subcapsular Cataracts Using IOL Master 700 and CASIA-2.. American journal of ophthalmology. 2023. DOI: 10.1016/j.ajo.2023.06.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.