Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2012

First inappropriate implantable cardioverter defibrillator therapy is often due to inaccurate device programming: analysis of the French OPERA registry.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Leenhardt Antoine, Defaye Pascal, Mouton Elisabeth, Delay Marc, Delarche Nicolas, Dupuis Jean-Marc, Bizeau Olivier, Mabo Philippe, Cheggour Saida, Babuty Dominique, OPERA Registry Investigators

Аннотация

AIMS: Inappropriate therapy delivered by implantable cardioverter defibrillators (ICDs) remains a challenge. The OPERA registry measured the times to, and studied the determinants of, first appropriate (FAT) and inappropriate (FIT) therapies delivered by single-, dual- and triple-chamber [cardiac resynchronization therapy defibrillator (CRT-D)] ICD.

METHODS AND RESULTS: We entered 636 patients (mean age = 62.0 ± 13.5 years; 88% men) in the registry, of whom 251 received single-, 238 dual-, and 147 triple-chamber ICD, for primary (30.5%) or secondary (69.5%) indications. We measured times to FAT and FIT as a function of multiple clinical characteristics, examined the effects of various algorithm components on the likelihood of FAT and FIT delivery, and searched for predictors of FAT and FIT. Over 22.8 ± 8.8 months of observation, 184 patients (28.9%) received FAT and 70 (11.0%) received FIT. Ventricular tachycardia (VT) was the trigger of 88% of FAT, and supraventricular tachycardia was the trigger of 91% of FIT. The median times to FIT (90 days; range 49-258) and FAT (171 days; 50-363) were similar. The rate of FAT was higher (P <0.001) in patients treated for secondary than primary indications, while that of FIT were similar in both groups. Out of 57 analysable FIT, 27 (47.4%) could have been prevented by fine tuning the device programming like the sustained rate duration or the VT discrimination algorithm.

CONCLUSIONS: First inappropriate therapy occurred in 11% of 636 ICD recipients followed for ∼2 years. Nearly 50% of FIT could have been prevented by improving device programming.