MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.hrthm.2023.01.024

Медицинская библиотека · Heart rhythm · 2023

Radiotherapy-induced malfunctions of cardiac implantable electronic devices: A meta-analysis.

Xu Beizheng, Wang Yueying, Tse Gary, Chen Jiayi, Li Guangping, Korantzopoulos Panagiotis, Liu Tong

Систематический обзорМетаанализСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Radiation therapy (RT) may pose acute and long-term risks for patients with cardiac implantable electronic devices (CIEDs), including pacemakers (PMs) and implantable cardioverter-defibrillators (ICDs).

OBJECTIVE

We conducted a systematic review and meta-analysis to examine the association between RT and PM/ICD malfunctions in patients with cancer.

METHODS

We searched the literature using the PubMed, the Cochrane Library the Web of Science, and Embase for relative publications until April 2022. Of the 550 initially identified studies, 17 retrospective observational studies including 2454 patients were finally analyzed.

RESULTS

The meta-analysis showed that RT was associated with an increased risk of ICD malfunctions (odds ratio [OR] 2.75; 95% confidence interval [CI] 1.74-4.33). Five studies were included in the subgroup analysis regarding photon beam energy, showing that radiation-induced CIED failure was more likely to occur in ICDs when beam energy was ≥10 MV (OR 5.28; 95% CI 2.14-13.03). Neutron-generating RT significantly increased the risk of CIED malfunctions (OR 3.97; 95% CI 1.70-9.26), especially the risk of reset (OR 5.79; 95% CI 2.37-14.12; P = .0001). We did not find significant differences in the risk of CIED failure between chest RT and other RT sites (OR 1.09; 95% CI 0.63-1.88).

CONCLUSION

Our meta-analysis suggests that ICDs are more likely to be affected by RT than PMs. These adverse events, especially reset, in patients with cancer were associated with neutron-generating RT and beam energy ≥10 MV. Given the increasing requirement for RT in several patients with cancer as well as the increasing implantation rates of CIEDs, a better risk stratification is needed in this setting.

Для цитирования: Xu Beizheng, Wang Yueying, Tse Gary, Chen Jiayi, Li Guangping, Korantzopoulos Panagiotis, Liu Tong. Radiotherapy-induced malfunctions of cardiac implantable electronic devices: A meta-analysis.. Heart rhythm. 2023. DOI: 10.1016/j.hrthm.2023.01.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.