DOI 10.1167/tvst.7.4.8

Медицинская библиотека · Translational vision science & technology · 2018

Safety and Feasibility of a Novel Sparse Optical Coherence Tomography Device for Patient-Delivered Retina Home Monitoring.

Maloca Peter, Hasler Pascal W, Barthelmes Daniel, Arnold Patrik, Matthias Mooser, Scholl Hendrik P N, Gerding Heinrich, Garweg Justus, Heeren Tjebo, Balaskas Konstantinos, de Carvalho J Emanuel Ramos, Egan Catherine, Tufail Adnan, Zweifel Sandrine A

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To study a novel and fast optical coherence tomography (OCT) device for home-based monitoring in age-related macular degeneration (AMD) in a small sample yielding sparse OCT (spOCT) data and to compare the device to a commercially available reference device.

METHODS

In this prospective study, both eyes of 31 participants with AMD were included. The subjects underwent scanning with an OCT prototype and a spectral-domain OCT to compare the accuracy of the central retinal thickness (CRT) measurements.

RESULTS

Sixty-two eyes in 31 participants (21 females and 10 males) were included. The mean age was 79.6 years (age range, 69-92 years). The mean difference in the CRT measurements between the devices was 4.52 μm (SD ± 20.0 μm; range, -65.6 to 41.5 μm). The inter- and intrarater reliability coefficients of the OCT prototype were both >0.95. The laser power delivered was <0.54 mW for spOCT and <1.4 mW for SDOCT. No adverse events were reported, and the visual acuity before and after the measurements was stable.

CONCLUSION

This study demonstrated the safety and feasibility of this home-based OCT monitoring under real-life conditions, and it provided evidence for the potential clinical benefit of the device.

TRANSLATIONAL RELEVANCE: The newly developed spOCT is a valid and readily available retina scanner. It could be applied as a portable self-measuring OCT system. Its use may facilitate the sustainable monitoring of chronic retinal diseases by providing easily accessible and continuous retinal monitoring.

Для цитирования: Maloca Peter, Hasler Pascal W, Barthelmes Daniel, Arnold Patrik, Matthias Mooser, Scholl Hendrik P N, Gerding Heinrich, Garweg Justus, Heeren Tjebo, Balaskas Konstantinos, de Carvalho J Emanuel Ramos, Egan Catherine, Tufail Adnan, Zweifel Sandrine A. Safety and Feasibility of a Novel Sparse Optical Coherence Tomography Device for Patient-Delivered Retina Home Monitoring.. Translational vision science & technology. 2018. DOI: 10.1167/tvst.7.4.8.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.