Micro-dislodgement of a self-expanding transcatheter heart valve: Incidence, predictors, and outcomes.

Kim Won-Keun, Renker Matthias, Dörr Oliver, Nef Holger, Fischer-Rasokat Ulrich, Choi Yeong-Hoon, Hamm Christian W, Charitos Efstratios

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Malpositioning of transcatheter heart valves increases the risk of procedural failure. For the ACURATE system, inadvertent movement of the prosthesis to a varying extent is sometimes observed upon full release, but the incidence, mechanisms, and clinical impact of such valve micro-dislodgement (VMD) are poorly understood. The aim of the present study was to assess the incidence, predictors, and clinical outcomes of VMD in an all-comers population that underwent transcatheter aortic valve implantation (TAVI) with the ACURATE neo2 prosthesis (NEO2).

METHODS

This was a retrospective analysis of 448 consecutive patients who underwent transfemoral TAVI with NEO2 at our institution. VMD was defined as displacement ≥2 mm between the initial position and immediately after valve release as measured on fluoroscopy at the non-coronary cusp. The initial valve position prior to step 2 was categorized using the radiopaque marker band (RMB) relative to the annular plane. In addition, further anatomical and procedural characteristics were assessed.

RESULTS

A total of 68 (15.2%) cases with VMD were identified. A larger cover index, higher RMB position, partial detachment of the lower crown, and severe parallax prior to deployment were independent predictors of VMD, whereas a position of the delivery system in the outer curvature was protective against VMD. Among patients with VMD, the rates of valvular malpositioning and thus technical failure (VARC-3) were higher, but mean transprosthetic gradients were lower.

CONCLUSIONS

VMD occurs in a notable proportion of transfemoral TAVI cases with NEO2 and is associated with more frequent technical failure of the procedure.

Для цитирования: Kim Won-Keun, Renker Matthias, Dörr Oliver, Nef Holger, Fischer-Rasokat Ulrich, Choi Yeong-Hoon, Hamm Christian W, Charitos Efstratios. Micro-dislodgement of a self-expanding transcatheter heart valve: Incidence, predictors, and outcomes.. International journal of cardiology. 2022. DOI: 10.1016/j.ijcard.2022.04.031.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.