Validity and reliability of a point-of-care nerve conduction device in diabetes patients.
Аннотация
AIMS/
INTRODUCTION
Although nerve conduction study (NCS) using a standard electromyography system (EMGS) is considered to be the gold standard in evaluating diabetic polyneuropathy, this examination requires expensive equipment and well-trained technicians. We aimed to validate a point-of-care device, NC-stat/DPNCheck™, that has been developed for widespread use of NCS in diabetic polyneuropathy.
MATERIALS AND
METHODS
Diabetes patients underwent two kinds of NCS: DPNCheck™ and electromyography system. Inter-/intrarater reliability of DPNCheck™ were also determined by the intraclass correlation coefficient.
RESULTS
A total of 57 patients were evaluated. The parameters of NCS between the two methods correlated well (r = 0.7734 for the sural nerve conduction velocity, r = 0.6155 for the amplitude of sural nerve action potential). The intraclass correlation coefficients were excellent (intrarater: the velocity 0.767, the amplitude 0.811; interrater: the velocity 0.974, the amplitude 0.834).
CONCLUSIONS
The point-of-care device has excellent reproducibility and good agreement with standard electromyography system. The device might be useful to evaluate diabetic polyneuropathy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.