Ablation of persistent atrial fibrillation using multielectrode catheters and duty-cycled radiofrequency energy.

Scharf Christoph, Boersma Lucas, Davies Wyn, Kanagaratnam Prapa, Peters Nicholas S, Paul Vince, Rowland Edward, Grace Andrew, Fynn Simon, Dang Lam, Oral Hakan, Morady Fred

Клиническое исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

The purpose of this study was to assess the efficacy and safety of a novel, multielectrode, duty-cycled radiofrequency ablation (RFA) system for long-standing persistent atrial fibrillation (AF).

BACKGROUND

RFA for persistent AF remains a lengthy and challenging procedure.

METHODS

In 5 European centers, 50 patients with long-standing persistent AF underwent RFA. A circular pulmonary vein (PV) ablation catheter was used for PV isolation. Complex fractionated atrial electrograms were targeted at the interatrial septum using a multiarray septal ablation catheter and in the left atrium using a multiarray ablation catheter.

RESULTS

During a mean total procedure time of 155 +/- 40 min, complete PV isolation and complex fractionated atrial electrogram ablation were achieved in all patients. In 50% of patients, redo ablation was performed using the same strategy and technology. There were no device-related adverse events. At 6 months, a 7-day Holter electrocardiogram showed >80% AF reduction in 40 of 50 patients (80%), and 32 of 50 (64%) were off antiarrhythmic drugs. At 20 +/- 4 months after the last procedure, 31 of 47 patients (66%) had a >80% reduction in AF burden, with 21 patients (45%) free of AF and off antiarrhythmic drugs.

CONCLUSIONS

This initial 50-patient multicenter study demonstrates a 80% short-term and 66% success rate at 20 months, with a low complication rate and a relatively short procedure time in patients with persistent AF using 3 anatomically specific multielectrode ablation catheters and low-energy duty-cycled radiofrequency energy.

Для цитирования: Scharf Christoph, Boersma Lucas, Davies Wyn, Kanagaratnam Prapa, Peters Nicholas S, Paul Vince, Rowland Edward, Grace Andrew, Fynn Simon, Dang Lam, Oral Hakan, Morady Fred. Ablation of persistent atrial fibrillation using multielectrode catheters and duty-cycled radiofrequency energy.. Journal of the American College of Cardiology. 2009. DOI: 10.1016/j.jacc.2009.07.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.