DOI 10.1089/lrb.2022.0038

Медицинская библиотека · Lymphatic research and biology · 2023

Test-Retest Reliability of a Protocol for Assessment of Local Tissue Water in the Head and Neck Area.

Arends Coralie R, van der Molen Lisette, van den Brekel Michiel W M, Stuiver Martijn M

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Background

Lymphedema measurement is vital to select appropriate treatment and monitor its progress. Quantifying lymphedema in the head and neck area is challenging. The use of tissue dielectric constant (TDC) measurements has shown promising results in other body areas. This study aims to determine the test-retest reliability of a TDC measurement protocol developed for the head and neck area. Methods and

Results

A detailed measurement protocol, including eight measurement points per side, was developed. Subsequently, the reliability of the protocol was tested in a sample of healthy participants (n = 50, 28 males). Using the LymphScanner (Delfin, Finland), participants were subjected to two measurement sessions. Each measurement point was measured three times per session. Test-retest reliability for each point was evaluated using intraclass correlation coefficients (ICCs) and standard errors of measurement (SEMs). Using the average of three measurements, reliability was good to excellent for all points (ICCs 0.81-0.95), with small measurement errors (SEMs 1.51-2.86). The reliability of a single measurement was moderate to excellent for all measurement points (ICCs 0.58-0.87), with larger, but still small, measurement errors (SEMs 1.65-3.39). When using single measurements, the lowest ICCs were found for the temporal (left 0.73 and right 0.67) and submandibular (left 0.58 and right 0.77) locations.

Conclusion

Measurements with the LymphScanner, taken according to the developed protocol, are reliable in healthy participants. We recommend using the average of three measurements to optimize reliability. The protocol is fit for further testing in patient populations and for determining normal values in a larger scale study with healthy subjects.

Для цитирования: Arends Coralie R, van der Molen Lisette, van den Brekel Michiel W M, Stuiver Martijn M. Test-Retest Reliability of a Protocol for Assessment of Local Tissue Water in the Head and Neck Area.. Lymphatic research and biology. 2023. DOI: 10.1089/lrb.2022.0038.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.