BACKGROUND

Vertical displacement of the residual limb within transtibial prosthetic socket, often known as "pistoning" or downward movement, may lead to skin breakdowns and ulcers. Downward movement is particularly difficult to self-manage for diabetic individuals living with amputation because of diminished sensation in the residual limb from peripheral neuropathy. Therefore, a customizable sensor at the distal end that can alert the users when high-risk downward movement and pistoning occurs is urgently needed.

OBJECTIVES

Presented herein for the first time is a lightweight, inexpensive sensing system inspired by soft robotics that can detect the occurrence and severity of downward movement at the distal end.

METHODS

The sensing system consists of a multilayered torus-shaped balloon, allowing easy integration with pin-lock socket systems. The design allows sensing of vertical displacement without imparting high reaction forces back onto the distal end. A benchtop compression tester was used to characterize system performance. Systematic and parametric benchtop tests were conducted to examine the sensor's physical characteristics. Long-term (24-h) stability of the sensor was also recorded.

RESULTS

Compared with water, air was determined to be a better medium with a higher linear full-scale span (FSS) because of its compressible nature. Repeatable 0.5-mm vertical displacements yielded a linear (>0.99 R 2 ) FSS of 4.5 mm and a sensitivity of 0.8 kPa/mm. The sensing system is highly precise, with as low as 1% FSS total error band and average hysteresis of 2.84% of FSS. Over 24 h, a 4% FSS drift was observed.

CONCLUSION

Sensing system characteristics, coupled with low-cost, customizable fabrication, indicates promising performance for daily use to notify and alert transtibial prosthetic users of downward movement and/or pistoning.

Для цитирования: Yu Adam J, Gao Run Ze, Lee Peter S, Mele Christian, Dittmer Doug, Schirm Andreas, Ren Carolyn L, Tung James Y. Soft robotics-inspired sensing system for detecting downward movement and pistoning in prosthetic sockets: A proof-of-concept study.. Prosthetics and orthotics international. 2023. DOI: 10.1097/pxr.0000000000000302.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.