Validation of Proprietary and Novel Step-counting Algorithms for Individuals Ambulating With a Lower Limb Prosthesis.

Rigot Stephanie K, Maronati Rachel, Lettenberger Ahalya, O'Brien Megan K, Alamdari Kayla, Hoppe-Ludwig Shenan, McGuire Matthew, Looft John M, Wacek Amber, Cave Juan, Sauerbrey Matthew, Jayaraman Arun

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare the accuracy and reliability of 10 different accelerometer-based step-counting algorithms for individuals with lower limb loss, accounting for different clinical characteristics and real-world activities.

DESIGN

Cross-sectional study.

SETTING

General community setting (ie, institutional research laboratory and community free-living).

PARTICIPANTS

Forty-eight individuals with a lower limb amputation (N=48) wore an ActiGraph (AG) wGT3x-BT accelerometer proximal to the foot of their prosthetic limb during labeled indoor/outdoor activities and community free-living.

INTERVENTIONS: Not applicable.

MAIN OUTCOME MEASURES: Intraclass correlation coefficient (ICC), absolute and root mean square error (RMSE), and Bland Altman plots were used to compare true (manual) step counts to estimated step counts from the proprietary AG Default algorithm and low frequency extension filter, as well as from 8 novel algorithms based on continuous wavelet transforms, fast Fourier transforms (FFTs), and peak detection.

RESULTS

All algorithms had excellent agreement with manual step counts (ICC>0.9). The AG Default and FFT algorithms had the highest overall error (RMSE=17.81 and 19.91 steps, respectively), widest limits of agreement, and highest error during outdoor and ramp ambulation. The AG Default algorithm also had among the highest error during indoor ambulation and stairs, while a FFT algorithm had the highest error during stationary tasks. Peak detection algorithms, especially those using pre-set parameters with a trial-specific component, had among the lowest error across all activities (RMSE=4.07-8.99 steps).

CONCLUSIONS

Because of its simplicity and accuracy across activities and clinical characteristics, we recommend the peak detection algorithm with set parameters to count steps using a prosthetic-worn AG among individuals with lower limb loss for clinical and research applications.

Для цитирования: Rigot Stephanie K, Maronati Rachel, Lettenberger Ahalya, O'Brien Megan K, Alamdari Kayla, Hoppe-Ludwig Shenan, McGuire Matthew, Looft John M, Wacek Amber, Cave Juan, Sauerbrey Matthew, Jayaraman Arun. Validation of Proprietary and Novel Step-counting Algorithms for Individuals Ambulating With a Lower Limb Prosthesis.. Archives of physical medicine and rehabilitation. 2023. DOI: 10.1016/j.apmr.2023.10.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.