DOI 10.1016/s0886-3350(97)80187-3

Медицинская библиотека · Journal of cataract and refractive surgery · 1997

Comparison of keratometric values of healthy and diseased eyes measured by Javal keratometer, EyeSys, and PAR.

Varssano D, Rapuano C J, Luchs J I

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the simulated keratometric results of the PAR CTS and the EyeSys corneal imaging systems with Javal keratometer readings in keratoconus eyes, in eyes after corneal grafting, and in healthy eyes.

SETTING

Cornea Service, Wills Eye Hospital, Philadelphia, Pennsylvania, USA.

METHODS

This prospective study evaluated 69 patients divided into three groups (keratoconus; 6 to 12 months postkeratoplasty; normal corneas) treated at the cornea service. The eyes were measured with each of the three instruments. Keratometry results and number of attempts needed to obtain the results were collected.

RESULTS

Thirty keratoconus, 18 transplanted, and 21 normal corneas were examined. The PAR CTS Imaged all corneas and the EyeSys, 86% (P = .000627). The PAR CTS and Javal flat and steep K-readings in all three groups were the same. The EyeSys simulated keratometry results were lower than those of Javal keratometer in the flat K (P < .00001) and steep K (P < .00001) in the normal group and in the steep K in the transplanted cornea group (P = .00823). The EyeSys also measured less astigmatism than the Javal in the normal (P = .00124) and transplanted cornea groups (P = .00376).

CONCLUSION

The PAR CTS was better able than the EyeSys to provide keratometric values that agreed with those obtained with the Javal keratometer.

Для цитирования: Varssano D, Rapuano C J, Luchs J I. Comparison of keratometric values of healthy and diseased eyes measured by Javal keratometer, EyeSys, and PAR.. Journal of cataract and refractive surgery. 1997. DOI: 10.1016/s0886-3350(97)80187-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.