DOI 10.1093/ejcts/ezab443

Медицинская библиотека · European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery · 2022

Transcatheter aortic valve replacement for structural degeneration of previously implanted transcatheter valves (TAVR-in-TAVR): a systematic review.

Gallo Michele, Fovino Luca Nai, Blitzer David, Doulamis Ilias P, Guariento Alvise, Salvador Loris, Tagliari Ana Paula, Ferrari Enrico

СтатьяСистематический обзорПеревод пока не подготовлен

Аннотация

OBJECTIVES

Transcatheter aortic valve replacement (TAVR) represents a valid treatment for patients with aortic valve stenosis and high or intermediate surgical risk. However, biological transcatheter valves can also experience a structural degeneration after years, and a redo-TAVR procedure (TAVR-in-TAVR) can be a valid option. We revised the current available literature for indications, procedural and technical details and outcome on TAVR-in-TAVR procedures for degenerated TAVR valves.

METHODS

A systematic search was conducted in the public medical database for scientific articles on TAVR-in-TAVR procedures for degenerated transcatheter valves. Data on demographics, indications, first and second transcatheter valve type and size, mortality, complications and follow-up were extracted and analysed.

RESULTS

A total of 13 studies (1 multicentre, 3 case series, 9 case reports) were included in this review, with a total amount of 160 patients treated with TAVR-in-TAVR procedures for transcatheter valve failure. The mean age was 74.8 ± 7.8 with 84 males (52.8%). The mean elapsed time from the first TAVR procedure was 58.1 ± 23.4 months. Main indication for TAVR-in-TAVR was pure stenosis (38.4%, with mean gradient of 44.5 ± 18.5 mmHg), regurgitation (31.4%), mixed stenosis and regurgitation (29.5%) and leaflet thrombosis (8.8%). Procedural success rate was 86.8%, with second TAVR valve malposition occurred in 4 cases (2.5%). The hospital mortality rate was 1.25% (2/160). Post-procedural echocardiographic control showed moderate regurgitation in 5.6% of patients (9/160) and residual transvalvular mean gradient ≥20 mmHg in 5% of cases. Postoperative complications included major vascular complications (8.7%), new pacemaker implantation (8.7%), acute kidney failure (3.7%), stroke (0.6%) and coronary obstruction (0.6%). The mean follow-up time was 6 ± 5.6 months with 1 non-cardiovascular death reported.

CONCLUSIONS

TAVR-in-TAVR represents a valid alternative to standard surgery for the treatment of degenerated transcatheter valves in high-risk patients. Despite these promising results, further studies are required to assess durability and haemodynamic performances of the second TAVR valve.

Для цитирования: Gallo Michele, Fovino Luca Nai, Blitzer David, Doulamis Ilias P, Guariento Alvise, Salvador Loris, Tagliari Ana Paula, Ferrari Enrico. Transcatheter aortic valve replacement for structural degeneration of previously implanted transcatheter valves (TAVR-in-TAVR): a systematic review.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2022. DOI: 10.1093/ejcts/ezab443.