Ultrasonic versus high-speed cavity preparation: analysis of increases in pulpal temperature and time to complete preparation.

Vanderlei Aleska Dias, Borges Alexandre Luiz Souto, Cavalcanti Bruno Neves, Rode Sigmar Mello

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

STATEMENT OF PROBLEM: The use of ultrasonic tips has become an alternative for cavity preparation. However, there are concerns about this type of device, particularly with respect to intrapulpal temperatures and cavity preparation time.

PURPOSE: The purpose of this study was to analyze pulpal temperature increases generated by an ultrasonic cavity preparation with chemical vapor deposition (CVD) tips, in comparison to preparation with a high-speed handpiece with a diamond rotary cutting instrument. The time required to complete the cavity preparation with each system was also evaluated.

MATERIAL AND

METHODS

Thermocouples were positioned in the pulp chamber of 20 extracted human third molars. Slot-type cavities (3 x 3 x 2 mm) were prepared on the buccal and the lingual surfaces of each tooth. The test groups were: high-speed cavity preparation with diamond rotary cutting instruments (n=20) and ultrasonic cavity preparation with CVD points (n=20). During cavity preparation, the increases in pulpal temperature, and the time required for the preparation, were recorded and analyzed by Student's t test for paired samples (alpha=.05).

RESULTS

The average pulpal temperature increases were 4.3 degrees C for the high-speed preparation and 3.8 degrees C for the ultrasonic preparation, which were statistically similar (P=.052). However, significant differences were found (P<.001) for the time expended (3.3 minutes for the high-speed bur and 13.77 minutes for the ultrasound device).

CONCLUSIONS

The intrapulpal temperatures produced during cavity preparation by ultrasonic tips versus high-speed bur preparation were similar. However, the use of the ultrasonic device required 4 times longer for the completion of a cavity preparation.

Для цитирования: Vanderlei Aleska Dias, Borges Alexandre Luiz Souto, Cavalcanti Bruno Neves, Rode Sigmar Mello. Ultrasonic versus high-speed cavity preparation: analysis of increases in pulpal temperature and time to complete preparation.. The Journal of prosthetic dentistry. 2008. DOI: 10.1016/s0022-3913(08)60158-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.