OBJECTIVE

To develop a newly designed double-fixed keratoprosthesis (Seoul-type keratoprosthesis [S-KPro]) and to assess its mechanical stability and biocompatibility.

METHODS

Twenty-five rabbits were divided into 4 groups by fixation technique, amniotic membrane (AM) implantation, and skirt material. The eyes were studied with the use of slitlamp, light, and electron microscopy. Stress testing was performed. In addition, 2 human subjects underwent S-KPro implantation. Best-corrected visual acuity was checked, and ophthalmic examination was performed.

RESULTS

The average retention period of the group receiving double-fixated polyurethane-S-KPro with AM was longer (>24 weeks) than that of the others. Fibroblast invasions were found in polyurethane pores but not in polytetrafluoroethylene (Gore-Tex) pores on light microscopy. The minimal pressure that induced aqueous leakage was greater than 250 mm Hg in all of the tested eyes. Two human subjects have maintained a good postoperative condition for 18 and 8 months.

CONCLUSIONS

The double-fixation technique of applied S-KPro and AM appears to be helpful in improving the stability of the keratoprosthesis. Polyurethane with relatively large pore size (40 microm) may be used successfully as a material for the keratoprosthesis skirt.

CLINICAL RELEVANCE: Our results may be important for improving the clinical outcome of keratoprosthesis.

Для цитирования: Lee J H, Wee W R, Chung E S, Kim H Y, Park S H, Kim Y H. Development of a newly designed double-fixed Seoul-type keratoprosthesis.. Archives of ophthalmology (Chicago, Ill. : 1960). 2000. DOI: 10.1001/archopht.118.12.1673.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.