DOI 10.1097/j.jcrs.0000000000000909

Медицинская библиотека · Journal of cataract and refractive surgery · 2022

Repeatability of a Scheimpflug tonometer to measure biomechanical parameters before and after myopic refractive surgery.

Lu Nan-Ji, Hafezi Farhad, Rozema Jos J, Hillen Mark, Hafezi Nikki, Zhang Jia, Koppen Carina

СтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the repeatability of several corneal biomechanical parameters with a Scheimpflug tonometer (Corvis ST) in myopic eyes and eyes that underwent transepithelial photorefractive keratectomy (transPRK), small-incision lenticule extraction (SMILE), or femtosecond laser-assisted in situ keratomileusis (FS-LASIK) surgery.

SETTING

Eye Hospital of Wenzhou Medical University, Wenzhou, China.

DESIGN

Prospective randomized controlled study.

METHODS

315 eyes from 315 patients (135 myopes, 58 post-transPRK, 52 post-SMILE, and 70 post-FS-LASIK) were included. 3 consecutive scans were performed to evaluate the repeatability of the 40 parameters examined.

RESULTS

315 eyes were included. In all eyes, the coefficient of variation (CoV) for intraocular pressure (IOP) and biomechanical-corrected IOP (bIOP) ranged from 7.29% to 9.47% and 6.11% to 7.75%, respectively; the CoV of pachymetry was <0.8%. The intraclass correlation coefficient of Corvis Biomechanical Index-Laser Vision Correction (LVC) was 0.680 for post-transPRK, 0.978 for post-SMILE, and 0.911 for post-FS-LASIK. The CoV of Stress-Strain Index (SSI) was 204.93% for post-transPRK, 91.92% for post-SMILE, and 171.72% for post-FS-LASIK. The CoV of the 6 clinically important dynamic corneal response parameters ranged from 2.0% to 7.8% for myopia, 1.8% to 11.1% for post-transPRK, 2.1% to 8.7% for post-SMILE, and 1.8% to 8.8% for post-FS-LASIK.

CONCLUSIONS

Excellent intrameasurement repeatability of IOP, bIOP, and pachymetry was observed in all groups; SSI measurement in post-LVC corneas displayed more variation. Caution is warranted when assessing SSI in post-LVC corneas for the purpose of diagnosing iatrogenic ectasia or evaluating biomechanical remodeling of postoperative refractive corneas.

Для цитирования: Lu Nan-Ji, Hafezi Farhad, Rozema Jos J, Hillen Mark, Hafezi Nikki, Zhang Jia, Koppen Carina. Repeatability of a Scheimpflug tonometer to measure biomechanical parameters before and after myopic refractive surgery.. Journal of cataract and refractive surgery. 2022. DOI: 10.1097/j.jcrs.0000000000000909.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.