Journal of refractive surgery (Thorofare, N.J. : 1995) · 2009

Ocular aberrations measured by the Fourier-based WaveScan and Zernike-based LADARWave Hartmann-Shack aberrometers.

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Knapp Stefanie, Awwad Shady T, Ghali Claudine, McCulley James P

Аннотация

PURPOSE: To evaluate agreements in lower and higher order aberration measurements by two Hartmann-Shack wavefront-sensing devices.

METHODS: Using the VISX WaveScan and Alcon LADARWave aberrometers, ocular aberrations at a fixed optical zone of 6 mm were measured on 36 eyes of 18 patients. A tunable light intensity source was used to control pupil size, which was checked using infrared pupillometry. Repeatability of measurements was evaluated using the intra-class correlation coefficient with 3 consecutive measurements on each aberrometer.

RESULTS: Mean absolute defocus for WaveScan and LADARWave was 2.82 +/- 2.69 and 2.93 +/- 3.24 root-mean-square (RMS) microm, whereas astigmatism was 0.81 +/- 0.49 and 0.87 +/- 0.57 microm, respectively. Pearson correlation coefficients between the two aberrometers were 0.908 and 0.870 for defocus and astigmatism, respectively, whereas higher order aberration correlation was less tight (Pearson correlation coefficient=0.596 for coma, 0.746 for trefoil, 0.836 for spherical aberration, 0.637 for secondary astigmatism, and 0.963 for quadra-foil [P<.001 for all]). The LADARWave had a tendency to display more spherical aberration than the WaveScan, especially at high aberration values, with mean absolute difference in measurement of 0.12 +/- 0.08 microm, and only 44% of eyes having less than +/-0.10 RMS microm of difference. The mean total higher order aberration absolute difference was 0.14 +/- 0.14 microm, with only 50% of eyes within +/-0.1 RMS of agreement. Vector analysis revealed appreciable discrepancies in third- and fourth-order directional Zemike components, while showing similar values for fifth-order components. Intra-class correlation coefficient values for both aberrometers over different aberration orders showed excellent repeatability.

CONCLUSIONS: The WaveScan and LADARWave share similar lower order aberration measurements, but display significantly different higher order aberration values.

Предметные рубрики MeSH