m+Medelevator.RU
← Поиск публикаций
Pacing and clinical electrophysiology : PACE · 1994

The value of rate regularity and multiplicity measures to detect ventricular tachycardia in the presence of atrial fibrillation or flutter.

СтатьяПоддержка исследования: вне правительства СШАПоддержка правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Chiang C M, Jenkins J M, DiCarlo L A

Аннотация

The predominant cause of inappropriate therapy by implantable antitachycardia devices with pacing and nonpacing cardioverter defibrillators, is mistaking a fast ventricular response during atrial fibrillation or flutter with true ventricular tachycardia (VT). The distinction between these arrhythmias is an important consideration in addressing the problem of reducing false-positives in detection mechanisms for implantable devices. Dual chamber analysis that examines atrial and ventricular event ratios has been proposed as a solution to this problem, but would still fail in distinguishing paroxysmal VT requiring treatment from a fast but otherwise benign ventricular response during atrial fibrillation or flutter. In this study, two methods for discriminating these tachyarrhythmias were evaluated. Method 1 examined ventricular rate and rate regularity as a method for VT detection. Method 2 combined rate and regularity as well as an additional multiplicity criterion for recognition of atrial flutter with a fast ventricular response. In 20 patients, Method 1 had 100% sensitivity of VT detection and 80% specificity for detection of atrial fibrillation or flutter. Method 2 had 90% sensitivity and 90% specificity. These results suggest that use of these algorithms in future implantable devices would result in a decrease in false-positive device therapies.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты