MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcrs.2008.08.015

Медицинская библиотека · Journal of cataract and refractive surgery · 2008

Residual higher-order aberrations caused by clinically measured cyclotorsional misalignment or decentration during wavefront-guided excimer laser corneal ablation.

Wang Li, Koch Douglas D

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the impact of residual wavefront aberrations on optical image quality induced by cyclotorsional misalignment and centration error.

SETTING

Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA.

METHODS

Cyclotorsional misalignment and decentration were measured in 58 eyes of 38 patients using an iris-registration system. To calculate residual wavefront aberrations if iris registration was not engaged, each eye's preoperative wavefront map was rotated or decentered by the amounts calculated by iris registration; this wavefront was subtracted from the preoperative wavefront (6.0 mm pupil). The impact of the residual wavefront aberrations on optical quality were analyzed by (1) the root mean square of the total aberrations (2nd to 6th orders), lower-order aberrations (LOAs) (2nd order), and higher-order aberrations (HOAs) (3rd to 6th orders); (2) normalized polar Zernike coefficients; (3) polychromatic modulation transfer function (PMTF) with the Stiles-Crawford effect.

RESULTS

On average, centration error induced 4.9 times, 2.8 times, and 8.7 times higher values of total aberrations, LOAs, and HOAs, respectively, than cyclotorsional misalignment. For all Zernike terms except 2nd-order astigmatism, centration error produced significantly higher values than cyclorotational misalignment (all P<.05 with Bonferroni correction). Compared with the PMTF value with perfect alignment, cyclotorsional misalignment decreased the PMTF values slightly whereas centration error reduced the PMTF values by 31% to 66%.

CONCLUSION

Theoretically, image quality was reduced more by clinically measured decentration than by cyclotorsional misalignment.

Для цитирования: Wang Li, Koch Douglas D. Residual higher-order aberrations caused by clinically measured cyclotorsional misalignment or decentration during wavefront-guided excimer laser corneal ablation.. Journal of cataract and refractive surgery. 2008. DOI: 10.1016/j.jcrs.2008.08.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.