Monte Carlo simulation of expected outcomes with the AcrySof toric intraocular lens.
Аннотация
BACKGROUND
To use a Monte Carlo simulation to predict postoperative results with the AcrySof Toric lens, evaluating the likelihood of over- or under-correction using various toric lens selection criteria.
METHODS
Keratometric data were obtained from a large patient population with preoperative corneal astigmatism <= 2.50D (2,000 eyes). The probability distributions for toric marking accuracy, surgically induced astigmatism and lens rotation were estimated using available data. Anticipated residual astigmatism was calculated using a Monte Carlo simulation under two different lens selection scenarios.
RESULTS
This simulation demonstrated that random errors in alignment, surgically induced astigmatism and lens rotation slightly reduced the overall effect of the toric lens. Residual astigmatism was statistically significantly higher under the simulation of surgery relative to an exact calculation (p < 0.05). The simulation also demonstrated that more aggressive lens selection criteria could produce clinically significant reductions in residual astigmatism in a high percentage of patients.
CONCLUSION
Monte Carlo simulation suggests that surgical variability and lens orientation/rotation variability may combine to produce small reductions in the correction achieved with the AcrySof Toric IOL. Adopting more aggressive lens selection criteria may yield significantly lower residual astigmatism values for many patients, with negligible overcorrections. Surgeons are encouraged to evaluate their AcrySof Toric outcomes to determine if they should modify their individual lens selection criteria, or their default surgically induced astigmatism value, to benefit their patients.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.