Use of a noninvasive continuous monitoring device in the management of atrial fibrillation: a pilot study.

Rosenberg Michael A, Samuel Michelle, Thosani Amit, Zimetbaum Peter J

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Outpatient ambulatory cardiac rhythm monitoring is a routine part of the management of patients with paroxysmal atrial fibrillation (AF). Current systems are limited by patient convenience and practicality.

METHODS

We compared the Zio(®) Patch, a single-use, noninvasive waterproof long-term continuous monitoring patch, with a 24-hour Holter monitor in 74 consecutive patients with paroxysmal AF referred for Holter monitoring for detection of arrhythmias.

RESULTS

The Zio(®) Patch was well tolerated, with a mean monitoring period of 10.8 ± 2.8 days (range 4-14 days). Over a 24-hour period, there was excellent agreement between the Zio(®) Patch and Holter for identifying AF events and estimating AF burden. Although there was no difference in AF burden estimated by the Zio(®) Patch and the Holter monitor, AF events were identified in 18 additional individuals, and the documented pattern of AF (persistent or paroxysmal) changed in 21 patients after Zio(®) Patch monitoring. Other clinically relevant cardiac events recorded on the Zio(®) Patch after the first 24 hours of monitoring, including symptomatic ventricular pauses, prompted referrals for pacemaker placement or changes in medications. As a result of the findings from the Zio(®) Patch, 28.4% of patients had a change in their clinical management.

CONCLUSIONS

The Zio(®) Patch was well tolerated, and allowed significantly longer continuous monitoring than a Holter, resulting in an improvement in clinical accuracy, the detection of potentially malignant arrhythmias, and a meaningful change in clinical management. Further studies are necessary to examine the long-term impact of the use of the Zio(®) Patch in AF management.

Для цитирования: Rosenberg Michael A, Samuel Michelle, Thosani Amit, Zimetbaum Peter J. Use of a noninvasive continuous monitoring device in the management of atrial fibrillation: a pilot study.. Pacing and clinical electrophysiology : PACE. 2012. DOI: 10.1111/pace.12053.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.